64
To disable a latch feature, the corresponding asterisk must be removed.
Use the ENT button to accept and save new Latch Alarm Event Register
settings in the meter’s nonvolatile memory.
d) Reset Status Alarm Event Register (Tabular entry)
The
Status Alarm Event Register
can be reset by selecting the
“Reset Alarm Event
Reg”
menu option. A typical display with the
Status Alarm Event Register
reset
screen is shown in
Figure 37
. Note that it requires confirmation from the user:
Figure 37:
Resetting Alarm Events Register
Once the “YES” option is selected, the Event Register will be reset,
and the following confirmation screen will appear:
Figure 38:
Alarm Event Register Reset Confirmation Screen
***********************
Event Reg. Has
been reset!
***********************
Press any Key...
NOTE:
Any Alarm Events that may have occurred (Event 0 to Event
D) are stored in the internal status register. All detected events (if
corresponding bit in the latch register is not masked) remain stored
until the register is manually reset (by means of the digital
communication interface). If the event corresponding bit in the latch
register is masked (disabled), the event will be indicated as long as
it is active (no latching). The status Alarm Event Register is mapped
to the SRAM (volatile memory). In case of power interruption, the
status Event Register will be automatically reset.
Reset Alrm Events Reg.:
NO
YES
DO YOU WANT
RESET EVENT REG?