APPENDIX 4
SENDING COMMANDS TO THE DFC
RS485
The standard DFC comes with an RS485 interface. The protocol described below
allows for the unit using either a custom software program or a “dumb terminal”.
All values are sent as printable ASCII characters. The start character is always !and
the command string is terminated with a carriage return (line feeds are
automatically stripped out by the DFC:
!<Addr>, <Cmd>,Arg1,Arg2,Arg3,Arg4<CR>
WHERE:
!
Start character
Addr
RS485 device address in the ASCII representation of
hexadecimal (00 through FF are valid).**
Cmd
The one or two character command from the table above.
Arg1 to Arg4
The command arguments from the table above.
Multiple arguments are comma delimited.
CR
Carriage return character.
**
Default address for all units is 11.
Several examples of commands follow.
All assume that the DFC has been configured for address 15 (0F hex) on the RS485 bus:
1.
To put the unit in digital mode:
!0F,M,D<CR>
The DFC will reply:
!0FMD<CR>
2.
To set the flow of 50% of FS:
!0F,S,50.0<CR>
The DFC will reply:
!0FS50.0<CR>
3.
To get a flow reading:
!0F,F<CR>
The DFC will reply:
!0F50.0<CR>
(Assuming the flow is at 50% FS)
4.
Set the high alarm limit to 5% above Set point:
!0F,A,H,5.0<CR>
The DFC will reply:
!0FA5.0<CR>
31
Содержание DFC
Страница 8: ...4 FIGURE b 1 WIRING DIAGRAM FOR DFC TRANSDUCERS...
Страница 25: ...APPENDIX 1 COMPONENTS DIAGRAMS 21 DFC Digital PC Board Primary Side...
Страница 26: ...APPENDIX 1 CONTINUED 22 DFC Digital PC Board Secondary Side...
Страница 27: ...APPENDIX 1 CONTINUED 23 DFC Analog PC Board Primary Side...
Страница 28: ...APPENDIX 1 CONTINUED 24 DFC Analog PC Board Secondary Side...
Страница 36: ...32...
Страница 37: ...33...
Страница 38: ...34...
Страница 39: ...35...
Страница 40: ...36...