- 17 - - 17 -17
Rev D
17
Warning
– Lamp Temperature Exceeded
WARNING!
LAMP BOX TEMPERATURE
EXCEEDED AND LAMP
HAS BEEN TURNED OFF
This latching warning shuts
down the bulb because of a
ventilation problem which
could be a warm ambient
environment,
obstructed
vents, or a fan failure.
Assure proper ventilation
(clean vents and clearance).
Re-power the unit to clear
the
message.
Call
for
technical help if the problem
persists.
Warning
– Ballast Temperature
WARNING!
BALLAST BOX TEMP
HIGH, ENSURE PROPER
VENTILATION
This signals a ventilation
problem which could be a
warm ambient environment,
obstructed vents, or a fan
failure.
Assure
proper
ventilation (clean vents and
clearance). If the message
does not clear, call for
technical help.
Warning
– Ballast Temperature Exceeded
WARNING!
BALLAST BOX TEMP
EXCEEDED AND LAMP
HAS BEEN TURNED OFF
This latching warning shuts
down the bulb because of a
ventilation problem which
could be a warm ambient
environment,
obstructed
vents, or a fan failure.
Assure proper ventilation
(clean vents and clearance).
Re-power the unit to clear
the
message.
Call
for
technical help if the problem
persists.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com