Warning Messages
Warning messages display important information about the unit. The
following section describes the protocol for each message. Some
warnings automatically reset, while others require user intervention.
Visual warnings will always be accompanied with audio beeping.
Warning
– Power Supply Failure
WARNING!
POWER SUPPLY FAILURE
REFER TO
USER MANUAL
This warning indicates a
power supply failure. The
lamp will be powered down
and remain off. Cycle the
AC power. If the problem
persists, call for technical
help.
Warning
– Lamp Temperature
WARNING!
LAMP BOX TEMPERATURE
HIGH, ENSURE PROPER
VENTILATION
This warning signals a
ventilation problem which
could be a warm ambient
environment,
obstructed
vents, or a fan failure.
Assure proper ventilation
(clean vents and clearance).
The warning will clear if the
internal temperature of the
unit decreases below the
warning temperature. If the
message persists, call for
technical help.