7
Equipment Set-Up and Installation
Equipment Location
Prior to installing the ScaleGard
TM
HP Reverse Osmosis System, carefully plan the installation location for all system components.
Position the ScaleGard HP reverse osmosis base unit and the pressure tank in desired locations as per (Figure 1).
1. Access to feedwater, drain line connections and electrical outlet is required.
2. The ScaleGard HP System must be accessible and have at least 3 inches (7.62 cm) clearance on all sides of the system to facilitate servicing.
An additional 10" will be needed on both sides of the ScaleGard HP system to allow for the future installation of the SGHP Pre-EXP or the
SGHP RO-EXP manifolds.
3. The pressure tank should be installed within 15 ft (4.6 m) of the ScaleGard HP reverse osmosis base unit and 0.5” or larger tubing diameter is
required to complete installation.
Figure 1 — Plumbing Diagram
Bypass (Provided by Installer)
Check Valve (Provided by Installer)
1/2” NPT Feed
Water Inlet
HF90-CL-RO (Pre-Filter)
HFRO 500 (RO Membrane)
3/8” JG Outlet to drain
Pre-Filter sample valve
Install kit shutoff valve
(Sold Separately)
Install kit shutoff valve
(Sold Separately)
1/2” JG Outlet to tank
1/2” NPT Outlet to Equipment
Install kit Tee Fitting and
sample valve (Sold Separately)
1/2” Tank shut off valve (Comes with Tank)
RO sample valve
BEV140/BEV160
Post Filter
(Required/Sold Separately)
To Equipment
Legend
Provided by Installer
Installation Kit (PN-6841441)
Pressure Tank
(Required/Sold Separately)
Pressure Relieve Valve
(Sold Separately)
NOTICE
NOTICE
See safety information regarding reducing the risk associated with property damage due to water leakage or fl ooding on page 4 prior
to set-up and installation of the system.
Содержание ScaleGard HP
Страница 19: ...19 Notes...