Specifications
phySiCAl
ChARACteRiStiCS
Warming
unit
4.5 in. (11 cm) high x 7.5 in. (19 cm) wide x 10 in. (25 cm) long; wt.: 7 lb. 7 oz. (3.4 kg)
certiFicationS
IEC/EN 60601-1; CAN/CSA-C22.2, No. 601.1
claSSiFication
MEDICAL — GENERAL MEDICAL EQUIPMENT AS TO ELECTRICAL SHOCK, FIRE AND
MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601-1; CAN/CSA-C22.2,
No. 601.1; ANSI/AAMI ES60601-1:2005 CSA-C22.2 No. 60601-1:08; Control No. 4HZ8
Classified under IEC 60601-1 Guidelines (and other national versions of the Guidelines) as Class I,
Type B, Ordinary equipment, Continuous operation. Classified by Underwriters Laboratories Inc. with
respect to electric shock, fire and mechanical hazards only, in accordance with IEC/EN
2
60601-1 and in
accordance with Canadian/CSA C22.2, No. 601.1. Classified under the Medical Device Directive as a
Class IIb device.
leAkAge
CURRent
Meets leakage current requirements in accordance with UL/IEC 60601-1.
enviRonmentAl
ConditionS
operating
temperature
range
15° to 40°C (59° to 104°F)
Storage
temperature
range
-20° to 45°C (-4° to 113°F)
operating
humidity
10 to 85% RH, noncondensing
atmoSpheric
preSSure
range
50 kPa to 106 kPa
eleCtRiCAl
ChARACteRiStiCS
tempeRAtURe
ChARACteRiStiCS
device
rating
100-120 VAC, 50/60 Hz
220-240 VAC, 50/60 Hz
Set
point
temperature
41°C
maximum
heating
poWer
900 W
over
-
temperature
alert
Model 245:
43°C
Model 247:
48°C
FuSe
2 x T10A-H (250V) for 100-120 VAC
2 x T6.3A-H (250V) for 220-240 VAC
under
-
temperature
alert
33°C
FuSe
type
Time delay, high breaking
over
-
temperature
cutoFF
Model 245:
44°C (warming units Rev N and newer)
46°C (warming units Rev A to M)
Model 247:
50°C