3M
TM
Ranger
TM
Blood/Fluid and Irrigation Warming Systems
English
12
Check the 3M
TM
Ranger
TM
system website to ensure you have the most recent version of this document.
www.rangerfluidwarming.com reorder #202524A
Blood/Fluid Warming System
Preventative Maintenance Checklist
caution
Temperature adjustments cannot be made to the Ranger warming unit. If the test shows
the temperature is out of specification, please call 3M Patient Warming technical support at
1-800-733-7775.
t
oolS
and
equipment
• Thermocouple: 0.005" lead or smaller (part #90020)
• Calibrated thermocouple temperature monitor
• Aluminum tape
• Ranger standard flow blood/fluid warming set
• Phillips screwdriver
• Saline/water
0.75"
1.80"
Thermocouple
Inlet tube
Aluminum tape
Drain 2mL or more
from cassette to remo
ve.
Insert cassette
before priming.
Fig. 1: Ranger standard flow blood/fluid warming cassette
M
od
el
2
45