method
1. Unplug the warming unit.
2. Unfold the cleaning tool. Wet the foam
pads with a nonabrasive solution, such as
Alconox
TM
brand detergent.
3. Insert the tool from the back of the unit
and pull the tool all the way out from the
front.
4. Rinse the tool with water and repeat 3 times. Discard the tool according to institutional protocol.
5. Wipe off the unit to remove excess fluid.
to
clean
reSiStant
,
dried
-
on
FluidS
:
1. Spray a nonabrasive solution inside the slot of the warming unit and let sit for 15-20 minutes.
2. Clean the unit by using the cleaning tool.
NOTE: You may use a nonmetal instrument, such as a cotton swab, to clean the upper channels. If you
are unable to adequately clean the unit, call 3M Patient Warming technical support.
S
toRAge
Store all components in a cool, dry place when not in use.
Call 3M Patient Warming technical support at 800-733-7775 or 952-947-1200 for service
information.