Operating QUESTempº 34 & 36
10
•
Flow
: On, Off .
•
Index: none, ACGIH, Navy, EPRI and select either: WBGTi (indoor),
WBGTo (outdoor) for Index setting.
Index Settings
Explanation
TLV and action limit
These index settings only apply to the
ACGIH Index. EPRI and Navy will ignore
this setting.
Clothing Correction
Parameters are set from 0 - 9.9
°
C. This is
a clothing correction for the WBGT in
degrees Celsius and is applied to the
selected WBGT when the work duration is
calculated. (It will not affect the WBGT as
displayed by the unit.) This value should
typically be set to 0.0 for the Navy. (The
field is noted as “Clo Corr”.)
3. Press the
I/O Enter
key to change a parameter. Time and date require
using the
Up/Down Arrows
and
I/O Enter
keys to modify each number.
NOTE: at any time, you can move back one level, by pressing the
Run/Stop key.
4. Exit Setup by pressing the
Run/Stop
key.
QTº34/36 allows printing to a parallel or serial printer or to a computer. The
QUESTempº 36 will recognize the cable plugged in and configure itself for serial or
parallel. If no cable is plugged in, it will default to serial. Press
I/O Enter
key to
begin printing. Press
Run/Stop
key to return to the menu.
NOTE: if you wish to stop the printing, press I/O enter key until you return to the
main menu. When the printer has stopped printing, remove the cable from the
printer to the instrument.
Reset
Resetting enables you to clear the logged data from memory. Press the
I/O Enter
key
to enter the
Reset mode
. Clear the memory by holding down the
I/O Enter
key
while the display counts down from three.