Getting Started
Turning on/off and basic operation
053-664, Rev.C
QUESTempº 44 & 46 models
4
Turning on/off and basic operation
To quickly get you started with the QUESTemp
º
44/46, the following section explains
how to turn on the instrument, run, and stop your session.
1.
Press the
I/O Enter
key to
turn on
. Proceeding the model and revision
information displayed on the screen, the main menu will appear.
Figure 1-3: Main menu of the QT
º
44/46
2.
Press the
I/O Enter
key (when view is selected) to access the measurement
screens.
(The Wet and Dry measurements screen will display. See Figure 1-6, page
10 for an example).
3.
To view different measurements, press the
Up
or
Down arrow
key to toggle
through the views.
NOTE:
There are 5-user selectable languages included in the QUESTemp
º
44/46. If you see the fields such as Wet, Dry, WBGTi, and WBGTo this indicates
the measurements are displaying in English.
To display an alternative language, select
Setup
from the main menu.
Press the
Down
arrow
repeatedly until “English” (or the appropriate
language) appears. Then repeatedly press the
I/O enter
key to toggle
through the languages. Once selected, all menus and measurement
screens will change to the selected language. To return to the main menu,
press the Run/Stop key.
4.
To return to the
main menu
, press and hold the
I/O Enter
key (3, 2, 1
countdown will appear) and the main menu will display.
a.
To select an option on the main menu, press the
up or down arrow
until an arrow appears directly in front of the appropriate menu
selection and then press
I/O enter
key.
5.
To
power off
, press and hold the
I/O enter key
from the main menu.
Placing the QUESTemp
º
44/46 on the job site
The QUESTemp
º
44/46 should be placed at a height of 3.5 feet (1.1m) for standing
individuals or 2 feet (.6m) for seated individuals. Tripod mounting is recommended to
get the unit away from anything that might block radiant heat or airflow. A 1/4"x 20
threaded bushing on the bottom of the instrument allows mounting to a standard
photographic tripod. Do not stand close to the unit during sampling.
Before datalogging, allow ten minutes for the sensors readings to stabilize.
Indicator arrow
Indicates the selected menu
option. Then press I/O enter to
select or press up/down arrows
to select another option
VIEW
SETUP
RESET