Exposure Index
Exposure Index is an approximate indicator of the dose that reaches the image receptor.
3DISC’s Exposure Index is calculated by following equation.
Under the same exposure conditions, the indicator and the dose are proportional. The
recommended EI value is between 1800 and 2200.
1800 < Dose < 2200: Dose is optimal
Dose < 1800: Dose is too low
Dose > 2200: Dose is too high