User Manual
QuantorMed Plus
TM -721-EN
47
7.2.
Exposure Input
Enter the X-ray exposure settings. Exposure Input field is only activated when it is enabled in
[System
– Exposure Input]. The entered values will be included in the DICOM information. As
shown in Table 7.1, 4 types can be entered.
Table 7.1. Exposure Input
Type
kVp
mA
ms
mAs
1
O
O
O
Calculated
automatically
2
O
X
X
O
3
O
O
Calculated
automatically
O
4
O
Calculated
automatically
O
O
7.3.
Cassette Indicator
The Cassette Indicator displays whether the cassette is inserted and recognized correctly.
Table 7.2 . Cassette Indicator
Indicator
Status
A cassette is not inserted.
A cassette is inserted, but cassette size information is not recognized.
A cassette is inserted, cassette size information is recognized, but the
current size has not been calibrated.
The status is okay and ready for scan.