User Manual
QuantorMed Plus
TM -721-EN
26
2.4.7.
Step 6: Scan High Dose
Figure 2.21.
Scan High Dose
The
Scan High Dose
button is enabled after performing a Scan Mid Dose.
Expose the cassette to the X-ray with the conditions noted below the
Scan High Dose
button.
Then, insert the cassette into the reader and click
Scan High Dose
button to acquire
the fourth calibration image.
The intensity value of the high dose image is displayed beneath the button.
The intensity value of the high dose is displayed in green if the value is within the
acceptable range.
The intensity value of the high dose is displayed in red if the value is not within the
acceptable range, click
Accept
to continue anyways or
Reject
to acquire the high dose
image again.
Increase the X-ray dose to increase the mean intensity value or lower the X-ray dose
to decrease the mean intensity value.
*If changes are made to technique (Dose)
please note the new values used (KVp, mAs) and input the new data in
C:\Qua\firecr.ini