PANNORAMIC® CONFOCAL RX 1.0
SCANNING MODES AND SETTINGS
MAY 13, 2021
63
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES .
Warning!
Do
NOT
use
Remove coverslip
function on TMA slides! The layout of TMA slides may mislead
the detecting algorithm.
Notice:
If a hematology slide is detected including blood smear, the
Remove Coverslip
function
cannot be activated and focusing on the coverslip area is not available. If you want to use
the
Remove Coverslip
function, first switch to
Custom Scanmap
.
Fill holes inside scanmap
–
If selected, the holes inside the specimen will also be scanned.
Often these holes contain small sample objects that otherwise would not be scanned.
Remove specks
–
If selected, the scanner does not digitize objects smaller than the
predetermined size. This helps filtering specks and other uninteresting objects out.
Set the maximum volume of specks manually in the setting window
.
Threshold settings
have two different statuses:
•
Auto threshold
–
It automatically detects the specimen in the preview
image.
Caution!
Make sure that the detection result is acceptable for you. Otherwise set the threshold value
manually.