PANNORAMIC® CONFOCAL RX 1.0
TROUBLESHOOTING AND MAINTENANCE
MAY 13, 2021
147
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES .
3.
Grab the pipette firmly at the top, then pull it outwards from its seating.
Warning!
Do
NOT
turn the pipette upside down, and make sure to hold it so that its tip points
downwards during the whole process of emptying and refilling.
Warning!
Pipette unit removal must be performed by an experienced and trained person only. If the
pipette unit is removed with not enough care, the skin of your hand can be bruised.
Figure 132: Type 2 pipette
4.
Rotate the knob anti-clockwise to empty the pipette entirely (
Type 1
)
OR
Remove the top cap of the pipette (
Type 2
).
5.
Dip the pipette head into the container for distilled water, and rotate knob clockwise until
endpoint to pump up the necessary amount of immersion fluid (
Type 1
)
OR
Hold the pipette vertically and top up with distilled water until its level reaches the
ma
rking „MAX LEVEL” (
Type 2
). Place cap back when finished.