background image

3D Systems, Inc.

3

AT A GLANCE ON GEOMAGIC CAPTURE APPLICATION

Scan

A

Live View

Displays current camera view.

B

Scanner Tab

Allows users to switch each scanner’s 

camera view.

C

Camera View Switch

Allows users to switch between the 

side of camera view.

D

Starting Position 

of 360º Rotational 

Scanning

Indicates the starting point of 

rotational scanning (from 0 to 360 

rotational degrees).

E

Scan

Starts the scanning.

F

Capturing Options

Options in this group allow users to 

manage their working environment 

and scan an object with optimal 

results.

G

Exposure Settings

Allows users to adjust an exposure 

value.

H

Turntable Settings

Allows users to adjust the number 

of scan divisions in 360 rotational 

degrees.

I

Status Bar

Shows the current working status of 

the scanner.

J

Settings

Goes to the Settings menu.

K

Help

Shows the Help document.

Settings

A

Device Status

Manages device connection status.

B

Exposure Settings

Allows users to adjust an exposure 

value.

C

Scan Resolution

Adjusts the resolutions of scanners.

D

Device Registration

Methods for registering the position of 

connected devices or aligning scans 

into one coordinate system

E

Advanced Options

Opens the Advanced Settings menu.

F

Back

Goes back to the Scan menu.

G

Help

Shows the Help document.

Содержание Geomagic Capture Series

Страница 1: ...Harness the Power of 3D Scanning Geomagic Capture User Guide Geomagic Capture Series 3D Scanner for 3D Scan Based Design and Quality Inspection...

Страница 2: ...CIFICATIONS 5 SYSTEM REQUIREMENTS FOR GEOMAGIC CAPTURE APPLICATION 5 4 3D SCANNING BASICS 6 5 SETTING UP GEOMAGIC CAPTURE SERIES 3D SCANNER 8 6 CREATING SCAN DATA 11 SCAN AN OBJECT AND CREATE SCAN DAT...

Страница 3: ...tion tools in a seamless push button manner With unprecedented performance and affordability Geomagic Capture ushers in a new era of design to manufacturing productivity and utility Harness the Power...

Страница 4: ...publication As more definitively stated and set forth in the License Agreement 3D Systems Inc does not and will not accept any financial or other responsibility that may result from use of the Softwar...

Страница 5: ...optimal results G Exposure Settings Allows users to adjust an exposure value H Turntable Settings Allows users to adjust the number of scan divisions in 360 rotational degrees I Status Bar Shows the...

Страница 6: ...Inc 4 Advanced Settings A Device Status Manages device connection status B Network Configuration Configures the network to communicate with scanners C Update Device Firmwares Updates device firmwares...

Страница 7: ...r End mm 124 x 120 mm Near 190 x 174 mm Far 87 x 68 mm Near 88 x 87 mm Far Calibration Pre calibrated Note It is not recommended that you use scanner setup mixed with different type of the Geomagic Ca...

Страница 8: ...e scanner so it can capture as much of the object as possible while remaining perpendicular to the object s surface Line of sight The scanner can capture only what is in its line of sight If you scan...

Страница 9: ...ts spray deodorant with titanium oxide and tempera paint from a compressor You must spray the part with a uniform light coat The thinner the coating the better Adding more than a thin coat of white po...

Страница 10: ...e check if all cables and cordsets are connected to the scanner and the PC and then click OK to reconnect the scanner 4 The Geomagic Capture application will initiate Select the check box next to the...

Страница 11: ...tion point in a network the IP address of the network adapter in your PC should be the same as a default gateway of the device and the subnet mask of the network adapter should be the same as the subn...

Страница 12: ...tarted up You can simply update the device firmware by clicking OK in the message Note If there is an issue with updating the device firmware you must use the manual device firmware update to control...

Страница 13: ...posure is set to the conditions Stability of the scanner and the object If either the scanner or the object move during the scan the data will be inaccurate Ensure both are stable and se cure Shape an...

Страница 14: ...stration methods for the scanners then click Register Note If a part has spherical features more than three for example you can register the position of the scanners into one coordinate system with th...

Страница 15: ...to the Scan menu by clicking Back 12 Reposition the part and click Scan to take another scan Note If a scan was made using the Best Fit alignment method make sure at least 30 of the original scan area...

Страница 16: ...elect the desired registration method then click Register The application will automatically take scans by rotating the plate in a certain degree then will calibrate the turntable by aligning the scan...

Страница 17: ...may have difficulty matching target markers and fail to align scans Please ensure the target markers are placed so cameras can see them even though the part is moved Note To align two scans at least...

Страница 18: ...d then go to Settings by clicking Settings to update the transforms to the application e In the Device Registration select Current Location then click Register to update the transforms that has been c...

Страница 19: ...part D Starting Position of Rotational Scanning Indicates the starting point of rotational scanning from 0 to 360 rotational degrees E Scanning Button Starts the scanning This button is available afte...

Страница 20: ...blue and black is under exposed An optimal scan has a good balance and several high spots of exposure Under Exposed Over Exposed Optimal Note Scan quality can vary by the following factors Environmen...

Страница 21: ...an exposure value A scan that shows a lot of red is over exposed and a scan that shows a lot of blue and black is under exposed This option is linked with the Exposure option in the Scan menu Under E...

Страница 22: ...s scanned from each connected scanner calculates a transform matrix so that the more than four scanned target markers can be matched and then uses the transforms for registering the position of the de...

Страница 23: ...3D Systems Inc 21 E Advanced Options Opens the Advanced Settings menu F Back Goes back to the Scan menu G Help Shows the Help document...

Страница 24: ...ice Status Auto Configure Automatically reconfigures the network for all checked scanners IP Address Used to manually set a static IP address Subnet Mask Default Gateway Used to manually set a subnet...

Страница 25: ...3D Systems Inc 333 Three D Systems Circle Rock Hill SC 29730 www 3dsystems com 2014 3D Systems Inc All rights reserved...

Отзывы: