2.4 Further limits
2.4.1 Environmental limits
The intended use presupposes the following environmental limits:
2.4.1.1 Transport
(systems packed correctly in original packaging)
Temperature range:
0°
–
40°C
Relative humidity:
max. 80% at temperatures up to 30°C,
then linear decrease to 50% at 40°C
Atmospheric pressure:
min. 700 hPa
2.4.1.2 Storage
(systems packed correctly in original packaging)
Temperature range:
0°
–
40°C
Relative humidity:
max. 80% at temperatures up to 30°C,
then linear decrease to 50% at 40°C
Atmospheric pressure:
min. 700 hPa
2.4.1.3 Use (normal mode / service mode)
•
In fixed structures
•
In rooms closed on all sides
•
In dry environments
•
In non-explosive atmospheres
•
Not in the vicinity of easily inflammable materials
•
On a stable flat surface of a suitable table with sufficient load-bearing capacity
and an ergonomically expedient working height (e.g. inspection workstation)
•
No direct sunlight
Temperature range:
10°
–
40°C
Relative humidity:
max. 80% at temperatures up to 30°C,
then linear decrease to 50% at 40°C
Altitude:
max. 2000 m above sea level
Atmospheric pressure:
min. 800 hPa
Mains voltage fluctuations:
max. 10% of nominal voltage
Cleanliness level
of the environment:
Cleanliness level SaS 1
(Clean zone according to VDA 19.2)
2.4.2 Additional devices / accessories required
•
No devices or equipment other than that supplied are required to operate the system.
•
Only accessories or devices that have been explicitly approved for the purpose in writing
by 3D Global Solutions GmbH may be used as optional accessories.
All possible uses not described in the intended use and its conditions, as well as modifications
in or to the device, are not permitted.