26
Last revised 16.02.12
Possible settings for the dose are:
1 J–240 J max.
Once all laser parameters have been set, this is shown in the display. The different values are now no longer flashing
and you can start with the laser treatment – see the chapter “Using the Laser.”
Please note:
The maximum energy dose rate that can be selected is determined by the maximum therapy time and the maximum
laser power level of the laser pen.
Example:
Maximum therapy time for the 3B LASER PEN 200 mW is 20 minutes and maximum laser power is 200 mW.
The maximum energy dose rate that can be set is therefore 200 mW * 20 minutes * 60 s = 240 J. The minimum dose
that can be selected is 1 J.
Note:
Please understand that due to the connection between energy and therapy time, changing one of the
two parameters will also impact the other parameter. The respective other value will be automatically adjusted
accordingly.
The selected value for the therapy time is confirmed and stored by pressing the SET key (3) again. You can now start
using the laser (see the chapter “Using the Laser”).
Please note that the settings for the parameters are retained when the unit is turned off and on again.
Using the Laser
The laser application can now be activated if the initial setting (display of the parameters) is shown on the screen.
200mW
A
240.0J
CW
Pressing the LASER key (1) activates the laser after a 2-sec. warm-up period. The LED (6) lights up green and the
display changes to the following screen:
100mW
M 19:43 - 1.7J
The display (5) now shows the remaining therapy time and the emitted energy dose rate in Joule as well as the
selected point search mode. The selected therapy time now counts backwards and the energy dose rate changes in
line with the selected laser power and the elapsed therapy time according to the formula “Energy dose rate (Joule)
= Laser power (mW) x Therapy time (sec.)”.
If the loudspeaker has been turned on, a signal will sound every 20 sec.
Treatment ends automatically after the selected therapy time has elapsed or when the selected energy dose has
been achieved. The end of the therapy is indicated by short triple sound signal.
Treatment can be terminated manually at any time by pressing the LASER key (1).