Luna G3 BPAP 25A (G3 B25A) User Manual MID-97/v1.0
6 / 48
Pressure Display Accuracy
±(0.8 cmH
2
O
+
4%)
Static Pressure Stability at 10 cmH
2
O
±0.5 cmH
2
O
Dynamic Pressure Stability
Pressures (cmH
2
O) 10BPM 15BPM 20BPM
6.5
±0.5
±0.5
±0.5
10
±1
±1
±1
20
±1
±1
±1
Device with humidification and 22mm Tube or Heated Tubing.
Ramp
The ramp time ranges from 0 to 60 minutes.
Sound Pressure Level
< 26 dB, when the device is working at the pressure of 10 cmH
2
O.
Sound Power Level
< 34 dB, when the device is working at the pressure of 10 cmH
2
O.
Maximum Flow
Test Pressures (cmH
2
O)
4
10
15
20
Measured Pressure at the
Patient Connection Port
(cmH
2
O)
3
9
14
19
Average Flow at the
Patient Connection Port
(L/min)
90
150
150
150
Air Tubing
Air tubing
Length
Inner diameter
Tubing
6 ft.(1.83m)
19mm
Heated Tubing
6 ft.(1.83m)
19mm
The Form and the Dimensions of the Patient Connection Port
The 22 mm conical air outlet complies with ISO 5356-1.
PM2.5 Filter
Efficiency:
>
90% for 2.5 micron dust