background image

 

MANUAL

 

MAMNET ECO 15 / 20 / 30 / 45 

Summary of Contents for MAMNET ECO 15

Page 1: ...MANUAL MAMNET ECO 15 20 30 45 ...

Page 2: ...y about it Before beginning installation read carefully this instructions manual and once the system has been started keep it in a safe place for future reference The instructions in this manual describe the operation of the models Mamnet Eco 15 Mamnet Eco 20 Mamnet Eco 30 Mamnet Eco 45 ORP and pH CHECK elements supplied Central Processing Unit CPU Electrode Electrode holder Manual Optional ORP 1 ...

Page 3: ...r supply cable before having made all power supply connections to the electrolysis cell Choose an installation site where the system will be easy to access and enable easy viewing of the CPU and the electrode TECHNICAL SPECIFICATIONS DIMENSIONS CPU SPECIFICATIONS CONTROL MICROPROCESSOR POWER SUPPLY 230 V ac 50 60Hz H50RR F 3G 3 x 1 mm2 OUTPUT 7 5V dc 15A MamNet Eco 15 6 8V dc 20A MamNet Eco 20 6 3...

Page 4: ...VC ADHESIVE Æ 63 NaCIO PRODUCTION 25ºC 4 5 gr L NaCl 15g h 360gr day 20gr h 480gr day 30g h 720gr day 45g h 1080gr day MINIMUM FLOW RECOMMENDED 7 53 mh ECO15 11 53 mh ECO20 163 mh ECO30 203 mh ECO45 No OF PLATES PER ELECTRODE 10 MAMNET ECO 45 FLOW DETECTOR 6 MAMNET ECO 30 FLOW DETECTOR 4 MAMNET ECO 20 FLOW DETECTOR 4 SHORT MAMNET 15 FLOW DETECTOR TEMPERATURE DETECTOR SEMI CONDUCTOR LOAD LOSS 0 15 ...

Page 5: ...ts in the technical area The CPUs STOP2 contact must be connected to a potential free contact in the in the purifier s control panel so that the pump and the MAMNET system are connected simultaneously If the pump stops for any reason the CPU will also stop ELECTRODE HOLDER The electrode holder is made of a transparent polymer which contains the electrode Ensure the electrode holder is in a place t...

Page 6: ...each model The float detector system is activated if there is no re circulation water flow through the cell or if that one is very low If the electrolysis gas is not evacuated a pouch will be generated that electronically isolates it from the auxiliary electrode Thus when the electrode is inserted into the electrode holder the level probe auxiliary electrode must be located on the upper part of th...

Page 7: ...No Flow or a large gas bubble may form CORRECT NOT ACCEPTABLE WARNING If the input and output valves to the pipe where the electrode holder is installed are closed simultaneously the flow detector will not work correctly with a risk of rupture resulting Although it is an extremely unusual situation that can be avoided by blocking once the equipment is installed the return valve to the swimming poo...

Page 8: ...n When Contact 1 is cathode Contact 2 is anode and vice versa IMPORTANT Due to the relatively high continual current intensity that circulates the electrode power supply cables their lengths and sections must not be modified without first consulting the authorised distributor The CPU ELECTRODE connection cable must be from the recommended section in this manual Temperature probe Auxiliary electrod...

Page 9: ...ate the equipment is in configu ration mode 8 Button to execute commands and move around the menu 9 Button to configure the system and move around the menu 10 Button for MAMNET ECO information and to move around the menu 11 LED to indicate whether the ORP function is connected or not 12 LEDto indicate that theelectrode needs to be checked 13 LED to indicate lack of water in the electrode holder 14...

Page 10: ...s changed by pressing CONFIG the SEAWATER screen will appear Use the buttons to change the current option no cleaning or every 3h 4h up to 8h The hours selected are those that will change the polarity The higher the calcium content in the water the less time will pass between changes in polarity Once the time has been selected by pressing CONFIG the next screen will appear and pressing ENTER will ...

Page 11: ...ET POINT HISTERESIS PROBE CALIBRATION CALIBRACION SONDA NO YES SETTING PROBE config YES config Yes CALIBRATE PROBE CALIBRAR SONDA CHANGE PH PA RAMS PROGRAMAR PH HIGH TARGET PH MAXIMO X XX LOW TARGET PH MINIMO X XX Yes PUMP RUN TIME SEGUNDOS BOMBEO XX SEG PUMP STOP TIME SEGUNDOS PARADA XX SEG Config No Yes PH REFERENCE 1 REFERENCIA PH 1 X XX PUT PROBE IN REFEREN CE SOLUT INTRODUCIR SONDA EN REFEREN...

Page 12: ...L and ppm from the last test made Pressing ENTER returns to the start screen Pressing INFO shows the screen where a new test can be made and Yes No can be used to select Selecting NO returns to the OPERATING HOURS screen Selecting YES starts a new test and after a few seconds the screen with the current reading will appear Pressing ENTER returns to the start screen Press INFO to return to the OPER...

Page 13: ... WORK HOURS HORAS ELECTRODO Partial total ORP type tipo CONCENTRATION CONCENTRACION xx g l xx ppm TEST wait espera CONCENTRATION CONCENTRACION xx g l xxx ppm TEMPERATURE SENSOR TEMPERATURA xx ºC xxx ºF NEW SALT TEST NUEVO TEST NO YES info info info info info info info Other key enter enter enter enter enter enter enter ...

Page 14: ...t is unusual a short circuit in the electrode will produce the same effect When the unit detects an overload it will go to Verification Mode and attempt to re start The unit will go into operation when it detects the problem is resolved It is recommended to verify the electrode and water conductivity status If necessary slightly dilute the salt concentration in the swimming pool While the alarm is...

Page 15: ... advise within another 500 hours RE POSITION When the electrode is inactive depleted the unit activates the alarm The saline electrolysis is made up of titanium laminates covered in high noble metal oxides This covering also has a limited lifetime After several thousand hours of service the covering will desactivate and must be replaced In order to confirm the depletion the output current must be ...

Page 16: ... treat with chlorine As an initial dose 2 gr m3 of trichloro isocyanuric powder or tablet maybe added to destroy the residual contamination of the salt Before beginning the work cycle disconnect the CPU and run the purifying pump for 24 hours to ensure the salt is completely dissolved During this time the intake will only occur through the drain that facilitates the salt dissolving process Next st...

Page 17: ... change the polarity This serves to loosen eventual small calcium deposits on the electrode There is an option to select not to make changes or to make changes every 1h up to 8h The selection will depend on the calcium level in the water cal h cal h Important note Chlorine production by means of electrolysis means that the pH will increase which means that calcium will form Also more heat means mo...

Page 18: ...o produce When the ORP value is above the set point electrolysis will stop The production value will vary depending on the set point difference and the actual value of the reading DIGITAL REGULATION When the user selects the ORP option with the CONFIG button production can be selected at 0 or 15 20 30 45gr L by programming the power free contact which may be NO or NC Opening or closing the contact...

Page 19: ...ystem is producing chlorine by performing the following measurements 1 Check that the filtration and the MAMNET ECO system are connected and operating 2 With an analysis test kit take a sample of water as close as possible to the swimming pool s driving nozzles To take the sample place the thumb over the end of the test tube and place it just in front of the nozzle to ensure the water sample taken...

Page 20: ...h pH value has a low HCIO Hypochlorous and a high CIO hypochlorite HCIO is actually the oxidant disinfectant agent in the swimming pool which explains that a pH of 6 8 7 2 is the most reasonable and recommended depending on the type of swimming pool In general when electrolysis system is used to control pH an agent must be applied solid liquid pH minus Products based on diluted sulphur are recomme...

Page 21: ... page 21 E Never scrape or brush the electrode Keep in mind that the electrode is made of titanium laminates covered in a layer of noble metal oxides Their use results in progressive wear and tear With the aim to optimise their lifetime keep the following aspects in mind 1 Do not clean the electrodes as described above too often 2 Although it is a SELF CLEANING saline electrolysis system prolonged...

Page 22: ...21 Max submergible limit ...

Page 23: ...ET POINT HISTERESIS PROBE CALIBRATION CALIBRACION SONDA NO YES SETTING PROBE config YES config Yes CALIBRATE PROBE CALIBRAR SONDA CHANGE PH PA RAMS PROGRAMAR PH HIGH TARGET PH MAXIMO X XX LOW TARGET PH MINIMO X XX Yes PUMP RUN TIME SEGUNDOS BOMBEO XX SEG PUMP STOP TIME SEGUNDOS PARADA XX SEG Config No Yes PH REFERENCE 1 REFERENCIA PH 1 X XX PUT PROBE IN REFEREN CE SOLUT INTRODUCIR SONDA EN REFEREN...

Page 24: ...23 ...

Page 25: ...t does at 100 power but when it reaches the set Hystereis point the power decreases progressively to achieve SET POINT ORP HIGH ALARM Value in mV above which an alarm is generated ORP LOW ALARM Value in mV below which an alarm is generated If the ORP probe breaks down there is a problem in the connection or some other fault occurs the values read may not be correct In these cases the Mvalues go to...

Page 26: ...25 ...

Page 27: ...he pool When water pH exceeds the set value the system will start automatically to pH minimum 2 pH MINIMUM Indicate the minimum pH value we want for our pool When the system is able to decrease the pH to this value it stops automatically 3 PUMP RUN TIME Indicate the time in which we operate dosing pump in seconds 4 PUMP STOP TIME Is a security system for pumping not to much acid in the pool becaus...

Page 28: ...warranty or until the expiration of the original warranty period This guarantee will only be valid when submitted the purchase invoice indicating the serial number of the product and it s date of purchase as well as your payment receipt The electrodes Active Titanium plates are elements that become wasted during the use due to consumption of the active layer in the process of electrolysis They are...

Page 29: ...28 NOTES ...

Page 30: ...29 ...

Page 31: ...30 ...

Page 32: ...31 ...

Reviews: