background image

USER MANUAL

CAREvent® ALS

Part Number: 17MP9656 - REV 16 Feb, 2021

AUTOMATIC AND MANUALLY 

TRIGGERED RESUSCITATOR 

01CV3000

Summary of Contents for 01CV3000

Page 1: ...USER MANUAL CAREvent ALS Part Number 17MP9656 REV 16 Feb 2021 AUTOMATIC AND MANUALLY TRIGGERED RESUSCITATOR 01CV3000 ...

Page 2: ...ctional Checks 8 2 1 Set Up 8 2 2 Testing of the Individual Features of the Ventilator 9 3 Operating Procedure 9 3 1 Connecting the Supply Hose and Patient Circuit 9 3 2 Manual Ventilation and Cardiac Compressions 10 3 3 Automatic Ventilation 11 3 4 Action to be Taken if Patient Vomits During Resuscitation 12 3 5 Demand Breathing and Automatic Circuit Shut Off 12 4 Servicing 13 4 1 Routine Mainten...

Page 3: ...ained in Cardio Pulmonary Resuscitation and the operation of oxygen powered ventilators should use this equipment Thoroughly review this instruction manual before use Note This resuscitator is intended for first response to a breathing emergency only and that patients must be transferred to a transport and emergency ventilator as equipment becomes available INTENDED USE The CAREvent CA resuscitato...

Page 4: ...ormal manner described in the instruction manual There are no other expressed or implied warranties This warranty gives no specific legal rights You may also have other rights that may vary according to local regulations 1 3 Features The CAREvent ALS Handheld Resuscitator is a pneumatically powered time volume cycled ventilatory resuscitator with the added feature of a Manually Actuated Automatic ...

Page 5: ...pre hospital care patient transport and for resuscitation and inter departmental transport in the hospital environment where the potential patient use is with children and adults Has a Manually Actuated Automatic Ventilation Override Button Manual Button Has a seven position selector providing six preset automatic settings for a range of patient sizes from children to adults with tidal volumes and...

Page 6: ...e 45 87 PSI 3 0 6 0 Bar Maximum Airway Pressure 60 cmH2 0 58 8 mBar Operating Temperature 18o C 50o C 0o F 122o F Storage Temperature 40o C 60o C 40o F 140o F Input Connection 9 16 DISS Patient Connection 15 22 mm Weight 16 OZ 0 45 Kg ALS 14 OZ 0 40 Kg BLS Size 140 x 63 x 73 mm 5 5 x 2 5 x 2 9 Cylinder Duration CAREvent ALS 31 103 minutes Cylinder Duration CAREvent BLS 69 138 minutes 1 4 Performan...

Page 7: ...ter use always ensure that all components are cleaned in accordance with the instructions provided in this manual 6 Ensure that all components are reassembled correctly and that all items are replaced in the carrying case 7 After use always ensure that a full air or oxygen cylinder is attached before returning the unit to its normal storage position 8 Ensure that a new sealing washer is used every...

Page 8: ...ilator The following features can be individually tested during the pre use Functional Check 1 Maximum Airway Pressure and Audible Alarm 2 Frequency Tidal Volume Adjustment 3 Manual Ventilation 4 Demand Valve Function and Automatic Circuit Shut Off 3 Operating Procedure 3 1 Connecting the Supply Hose and Patient Circuit The supply hose provided is attached to the oxygen inlet on the rear of the re...

Page 9: ...till further 1 If no respiratory effort is observed position yourself above the patient s head and apply the face mask over the patients mouth and nose The thumb and index fingers are used to hold the mask to the face while the remaining three fingers of each hand are placed along the angle of the jaw A tilt action is used to hyperextend the neck and move the jaw forward This helps displace the to...

Page 10: ...lish the correct tidal volume This is accomplished by moving the control toward the child setting if the blow off valve operates or towards the adult setting if chest rise is insufficient This can be done by simply using your thumb to slide the control without removing your hand from the mask WARNING Automatic ventilation of the patient who is intubated or whose mask is held in place with the opti...

Page 11: ...e resuscitator body being careful to ensure that the diaphragm b is retained Fig 3 3 Shake out any foreign material from the resuscitator diaphragm face mask and patient valve swivel housing 4 Rotate the selector to the appropriate position and either operate the manual button to blow out any contaminant or cycle automatically for a few breaths 5 Re assemble diaphragm patient valve and face mask a...

Page 12: ...his check is to ensure that all of the accessories and resuscitator components are present the air or oxygen cylinder is full and that the resuscitator is in working order Regulator working pressure suction if equipped and ventilator limiting pressures should be checked at least every six months and more frequently in high use applications Level I Preventative Maintenance should be performed annua...

Page 13: ...olution Under no circumstances should the complete unit be allowed to be soaked or immersed in cleaning solutions E Viewing Window D Manual Button C Setting Selector H Gas Supply Inlet G Resuscitator Body B Diaphragm A Patient Valve Swivel Housing F Pressure Relief System Disassembly of the resuscitator fig 3 DESCRIPTION PROCEDURE CRITERIA SCHEDULE BY PM Visual Inspection User Manual Chapter 4 1 M...

Page 14: ...if supplied using the extraction tool 5 Shake out any foreign material 6 Wash all components thoroughly in a mild soap solution and disinfect as required 7 The resuscitator can be wiped over with a soft cloth and mild soap solution 8 Dry all components thoroughly 9 Reassemble unit connect to an air or oxygen supply to check operation prior to packaging for emergency use 5 CAREvent ALS Handheld Res...

Page 15: ... In North America call Toll Free 1 800 387 3405 MedNet EC REP GmbH Borkstrasse 10 48163 Münster Germany EC REP SERIAL N0 www otwo com 45A Armthorpe Road Brampton ON Canada L6T 5M4 Telephone 1 905 792 OTWO 6896 N A Toll Free 1 800 387 3405 Facsimile 1 905 799 1339 Email resuscitation otwo com PHT DEHP ...

Reviews: