Universal Filter

 

Series

 

FT

®

 

 
 

FT-...-H2

 

 
Instruction Manual 
Version 1.00.01 

 
 
 
 
 
 
 
 

 

 
 
 
 
 
 

 
 

Summary of Contents for Universal Filter FT H2 Series

Page 1: ...Universal FilterSeriesFT FT H2 Instruction Manual Version 1 00 01 ...

Page 2: ... found in the annexe to this operating manual Please also contact our homepage www mc techgroup com for further information about our products There you can read or download the data sheets and operating manuals of all M C products as well as further information in German English and French This Operating Manual does not claim completeness and may be subject to technical modifications 04 2016 M C ...

Page 3: ...d application 7 6 1 Serial numbers 7 6 2 Power supply 7 6 3 Warnings and instructions 7 7 Technical Data 8 8 Description 9 9 Reception and storage 10 10 Preparation and Installation 10 11 Mounting 10 12 Electrical Connection 11 13 Starting 13 14 Maintenance 14 15 Spare parts list 15 16 Appendix 15 List of illustrations Figure 1 Flow characteristcs 9 Figure 2 Electrical connecting diagram 12 ...

Page 4: ...opriate transportation correct storage as well as professional installation and maintenance work All necessary information a skilled staff will need for appropriate use of this product are given in this operating manual 2 DECLARATION OF CONFORMITY CE Certification The product described in this operating manual complies with the following EU directives EMV Instruction The requirements of the EU dir...

Page 5: ...re opening the equipment it must be switched off and hold no voltages This also applies to any external control circuits that are connected The device is only to be used within the permitted range of temperatures and pressures Check that the location is weather protected It should not be subject to either direct rain or moisture The device must not be used in hazardous areas Installation maintenan...

Page 6: ...t minor physical injuries may occur in case the respective safety measures are not fulfilled C AR E Without the warning triangle means that a material damage may occur in case the respective safety measures are not met AT T E N T I O N This means that an unintentional situation or an unintentional status may occur in case the respective note is not respected NOTE These are important information ab...

Page 7: ...especially for hot gas analysis technique at 180 C 6 1 SERIAL NUMBERS The nameplates bearing the serial number are located on the mounting plate NOTE Always quote the device s serial number when making enquiries and ordering replacement parts 6 2 POWER SUPPLY The universal filter series FT H2 can be operated on alternating current in the range from 230 to 240V 50Hz or 115V 60Hz 6 3 WARNINGS AND IN...

Page 8: ...n classification electrical standard IP44 EN60529 EN61010 EN60519 1 Ambient temperature max 50 C Sample temperature max 180 C Ready for operation after approx 30 minutes Temperature alarm contact alarm point T 30 C at Tset contact rating 250V 3A 0 25A Regulation of temperature thermostat PT100 as option Sample connections G1 4 i DIN ISO 228 1 Stagnant space 65cm Filter surface 70cm Ma pressure 4 b...

Page 9: ... of the filter is shown on a thermometer For isolation and protection reasons the heated filter part is covered with a cap To prevent heating interruptions the in and outlet connection area is heated too The clamps are fixed in order to support the sample lines The filter in and outlet can be turned 180 on the oven top position to achieve easy mounting and flexible adaptation to local circumstance...

Page 10: ... cover of the heated filter after opening the two clamping devices Disconnect the thermal conductivity jaws on the left and right side of the filter head in by loosening the screws In order to connect the sample lines screw in suitably sized threaded connectors with a G connecting thread using PTFE sealing tape NOTE Use only connectors approved for 180 C Make sure that the connection is leak proof...

Page 11: ...high power electrical units with nominal voltages of up to 1000 V together with the associated standards and stipulations A main switch and matching fuse must be provided externally The main circuit must be equipped with a fuse corresponding to the nominal current over current protection for electrical details see technical data Remove the lid of the electrical connection box The electrical connec...

Page 12: ...12 FT H2 1 00 01 www mc techgroup com Figure 2 Electrical connecting diagram ...

Page 13: ... because of high surface temperatures Wear protective gloves and protect the filter from unauthorized access The following procedure is recommended Check the rated value setting on the built in thermostat NOTE If the rated value temperature needs to be lowered more than 30 C in one step duration operation the thermostat s excess temperature switch off is triggered for re start press reset key Swit...

Page 14: ...nditions a meaningful maintenance cycle must be elaborated for the specific application An indication that filter maintenance may be necessary could be shown by a constant decline in the amount of sample gas in the analysis system Filter maintenance is restricted essentially to replacing filter elements and checking seals The filter element has to be replaced as follows Remove cover lid Open the c...

Page 15: ... 3G Glas 3µm 75mm C a t r a t r a t r 90 F 0016 Filterelement F 0 1GF Glasfaser 0 1µm 75mm C a t r a t r a t r 90 F 0008 Filterelement F 3SS rostfr Stahl 3µm 75mm C a t r a t r a t r 90 F 0040 Viton O Ring R 2 4 8 90 F 3000 Heizpatrone HLPSR 100mm 230VAC 350W R 1 1 1 90 F 3010 Heizpatrone HLPSR 100mm 115VAC 350W R 1 1 1 90 P 5020 Thermostat 0 180 C für FT H2 R 1 1 1 90 F 0012 Filterglas F 120G R 1...