background image

User manual

Summary of Contents for FC-IR101

Page 1: ...User manual ...

Page 2: ...1 Operating Manual Infrared thermometer Model FC IR101 ...

Page 3: ...on definition 5 7 Function keys 6 8 Settings 7 9 Measuring method 7 9 1 Forehead temperature measurement 7 9 2 Ear temperature measurement 7 9 3 To be connected with your smart phone 8 10 Replacement of batteries 8 11 Cleaning and Maintenance 8 12 Trouble shooting 10 13 EMC Information 11 14 Commitment to quality and after sales service 14 15 Manufacturer information 14 ...

Page 4: ...om outdoor to indoor please put the thermometer and wait for about half an hour minutes before measuring After measuring extremely high or low temperature don t measure body temperatures immediately you should measure 10 minutes later Prohibit the thermometer into any liquid Don t use in high or low temperature environment for a long time No collision falling and sharp objects mix it is forbidden ...

Page 5: ...function Capacity of storing 35 sets measurements Switching between C and F Auto shut down and power saving Switching between mute and un mute mode Can be connected with smart phone via Bluetooth 4 0 4 Technical parameters Table 1 Measurement method Contact Type Measurement Mode Body Mode 32 0 C 42 9 C 89 6 F 109 2 F object Mode 0 C 100 C 32 F 212F Measurement accuracy Body Mode 0 2 C 0 4 F object...

Page 6: ...Weight About 70 g without battery Dimensions 160mm 39 mm 27 3mm length x width x height Product accessories 2x AAA batteries and 1x User Manual ...

Page 7: ...el 8 Battery cover 6 Display icon definition 1 Temperature value 2 Object temperature mode 3 Forehead temperature mode 4 Ear temperature mode 5 Battery level 6 Mute Un mute 7 Fahrenheit Celsius degrees 8 Bluetooth icon 9 Memory recall Table 2 Icon Definition Description Icon Definition Follow instructions for use General warning Caution Don t dispose with household waste after use ...

Page 8: ...n keys Table 3 Function key functional description Measure button Press to measure Memory button Press to get the latest 35 sets memories temp Shutdown button Keep pressing the button for 5 7 seconds device will power off Mute and un mute button Keep pressing the button for 2 3 seconds to switch from un mute to mute Conversion button Keep pressing for 5 7 seconds to switch from F to C Measure butt...

Page 9: ...hod 9 1 Forehead temperature measurement Press the Measure button to power on Position the thermometer on the forehead just above the middle of the eyebrow Press and release the Measure button in 1 second the beep is heard you can now read the value If the eyebrow area is covered with hair sweat or dirt please clean the area beforehand to improve the reading accuracy 9 2 Ear temperature measuremen...

Page 10: ... the measurement vale is between 37 5 37 9 it is considered low fever it is suggested to consult the doctor 9 3 To be connected with your smart phone 1 By searching nRF Connect from Android Market or App Store and download the software 2 Open the Bluetooth of your smart phone 3 Turn on your thermometer 4 Search Bluetooth in SCANNER of the software nRF Connect 5 Click CONNECT in the right side when...

Page 11: ...immersed in liquid contact your service personnel or service engineer 2 Before cleaning or disinfection make sure that the device is switched off and disconnected from the power line 3 After cleaning with isopropyl alcohol allow the thermometer to air dry Use a soft dry cloth to clean the thermometer body and LCD display 4 Do not use abrasive cleaners Never submerge this thermometer in water or an...

Page 12: ...ole unit low level disinfection is appropriate since it is a complete surface contacting device Clean the device before disinfection The validated disinfectant for cleaning the device is Isopropanol 70 Since Isopropanol is used for both cleaning and disinfecting a new cloth is required to be used for the disinfection step To disinfect the monitor follow these steps 1 Switch off the monitor and dis...

Page 13: ...ent Emissions test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The FC IR101 uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause and interference in nearby electronic equipment RF emissions CISPR 11 Class B The FC IR101 is suitable for use in all establishments other than domestic and those directly conn...

Page 14: ...nt Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth N A Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations in power supply input lines IEC 61000 4 11 5 UT 95 dip inUT for 0 5 cycle 40 UT 60 dip inUT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip inUT for 5 s N A Mains power quality shou...

Page 15: ...d with the following symbol NOTE 1 At 90MHz and 800MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM ...

Page 16: ...istance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80MHz and 800MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is aff...

Page 17: ...ever there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving antenna Increase the separation be...

Reviews: