background image

FV62 Rev. D   11/14/18

Page  3    

8. Adjust feature settings as desired. The sensor range is factory set. 

If adjustment is necessary, use a Phillips head screwdriver and turn 

GENTLY. SEE SENSOR RANGE ADJUSTMENT

. Replace electronics 

access cover and tighten securely with vandal-resistant hex wrench.

Before the supply water is turned on, be sure all stop valves to the fl ush 
valves are closed off tight. The stop valves can be opened and closed by 
using the adjusting screw located at the center of the stop valve cap, behind 
the adjusting screw cover if already installed. Stop valve adjustments can 
only be made by using the adjusting screw. 

It is not necessary to remove 

the stop valve cap when making adjustments. If for any reason it 
becomes necessary to remove the stop valve cap, be certain the water 
is shut off at the main supply valve.

9. When all fl ush valves are connected to the fi xtures and water pressure is 

available, it is recommended that the supply piping be fl ushed to remove 
dirt, pipe chips, etc., from system. This procedure should also be repeated 
when the system is drained for seasonal use, as occurs in athletic fi elds, 
recreation parks, etc.

10. The AquaSense “A” fl ush valves are preset for fi xture volume as 

marked on the valve cartons. The valve does not require regulation for 
variation in water pressure within its operating range. To set the fl ush 
valve for proper operation, open the stop valve completely by using the 
adjusting screw and fl ush the AquaSense “AV” several times. Gradually 
adjust the stop valve closed, using the adjusting screw, so that the 
rate of water fl ow into the fi xture is not excessive, yet is suffi cient  to 
adequately evacuate the waste. The fi nal setting for closets should 
provide a rate of water fl ow suffi cient enough to evacuate thirty sheets 
of toilet paper from the fi xture.  The  fi nal setting for urinals should be 
such that the fi xture will not overfl ow when the valve is actuated twenty 
times in succession. The stop cap screw cover should be replaced after 
fi nal adjustments have been made.

Sensor Range Adjustment

The range is factory set for urinal or closet applications. If necessary, the 
sensor range can be adjusted. Pressing the reset button will initiate a 
ten minute set-up period. During these ten minutes a red light will fl ash 
through the sensor lens when an object is in view of the sensor. Use this 
feature to properly set the sensor range. Slowly rotate the range adjustment 
clockwise to increase range and counterclockwise to decrease range. The 
range adjustment will stop at minimum and maximum settings. Do not turn 
beyond these stops. For an additional ten minutes of blinking, push reset 
button. Stand in front of sensor (minimum 10 seconds) and move forward 
and backward to determine range.

9

A. Remove the main valve 

body cover.

D. Place tube cap over 

riser tube. 

B. Remove the plastic 

cover.

C. Remove the diaphragm 

kit assembly from the 
fl ush valve

E. Replace the plas-

tic cover and main 
body cover without 
reinstalling the dia-
phragm kit assembly.

F. Open the stop valve by 

using the stop adjusting 
screw and flush out all 
debris from pipe and con-
nections.

G. Shut stop, open cover, remove tube cap, replace the 

diaphragm kit assembly, replace and tighten both covers. 

Option Setting
Switch 1 - Automatic Flush - The unit will automatically fl ush 24 hours after 

the last user. On position = activate, Off position = deactivate.

Switch 2 - Courtesy Flush - When an object has been detected for two 

seconds, the unit fl ushes then goes into normal mode. On position 
= activate, Off position = deactivate. (Always press reset button after 
changing switches.)

Switch 3 - Flush Time - Factory setting.
Switch 4 - Urinal Water Saver - Reduces fl ush volume by 40% during 

periods of continuous use. (Note: Always press reset button after 
changing switches.)

ON-OFF SWITCHES

RESET BUTTON

RANGE ADJUSTMENT

8

10

Sensor

Control Panel

Summary of Contents for AquaSense AV ZR Series

Page 1: ...provided we are notified in writing of such defect and the goods are returned to us prepaid at Sanford NC with evidence that they have been properly maintained and used in accordance with instructions We shall not be responsible for any labor charges or any loss injury or damages whatsoever including incidental or consequential damages The sole and exclusive remedy shall be limited to the replacem...

Page 2: ...nut with a wrench Do not turn water on until batteries are inserted see 8 2 Prior to inserting the flush valve tailpiece into stop valve be certain that the o ring seal is located in o ring seal groove at the end of the tailpiece and the locking nut and locking snap ring are located as shown Care should be taken not to damage the o ring when inserting the tailpiece into the stop valve If lubricati...

Page 3: ...he final setting for urinals should be such that the fixture will not overflow when the valve is actuated twenty times in succession The stop cap screw cover should be replaced after final adjustments have been made Sensor Range Adjustment The range is factory set for urinal or closet applications If necessary the sensor range can be adjusted Pressing the reset button will initiate a ten minute se...

Page 4: ...ent For the first ten minutes of operation the red light will flash when an object is detected 2 Low Battery Indication When the batteries have reached the caution level the red light will flash when an object is detected 3 Abnormal Reflection Indicator When a reflective surface is present for longer than 30 minutes the red light will flash indicating that an object is being continuously detected ...

Page 5: ...age while reassembling the unit Insert the solenoid Item 7 into the housing and install the solenoid retaining clip Item 8 When installed properly the end of the solenoid housing is free to rotate Note Be careful not to crimp the battery wires while inserting the solenoid into the housing 3 Reinstall the RetroFlush sensor assembly Item 9 onto the flush valve using a wrench Be sure the solenoid inl...

Page 6: ...Item 8 When installed properly the end of the solenoid is free to rotate Notes Be careful not to crimp the battery wires while inserting the solenoid into the housing Also be sure the blue and gray wires of the solenoid valve extend into the sensor module cavity of the housing Item 9 6 Connect the sensor module blue and gray wires to the solenoid valve blue and gray wires and reinstall the sensor ...

Page 7: ...lume Control 3 5 GpF 3 PR6000 EU25 WS Low Consumption Urinal Volume Control 1 6 GpF 3 PR6000 EU25 WS1 Diaphragm PR6000 E15A Sensor Module Replacement Kits Product No Solenoid Valve Replacement Kit 8 9 10 PR6000 M Sensor Lens Kit 12 PR6000 SC High Volume Sensor 13 PR6000 SH Low Volume Sensor 13 PR6000 SL Battery Holder with Connecting Cable 15 16 PR6000 MB Metal Housing 11 PR6000 MH Vandal Resistan...

Page 8: ...e or chemical cleaners to clean flush valve as they will dull the luster and attack the chrome or special decorative finishes Use only mild soap and water then wipe dry with a clean cloth or towel While cleaning the bathroom tile and floor the flush valve and electronics should be protected from splattering of water cleaner acids and cleaning fluids that can damage the sensor flush valve Problem C...

Reviews: