background image

Product Specifi cations

Sensor Range: 

12” to 60” (adjustable)

Battery Type :  

AA Duracell Alkaline – Model MN-1500

Battery Life: 

3 years at 4,000 fl ushes per month

Flush Volume: 

ZR-6000  

–  4.5 gallons per fl ush

 

ZR-6000-WS  

–  3.5 gallons per fl ush

 

ZR-6000-WS1   –  1.6 gallons per fl ush

 

ZR-6001  

–  3.0 gallons per fl ush

 

ZR-6003-WS  

–  1.5 gallons per fl ush

 

ZR-6003-WS1   –  1.0 gallons per fl ush

Voltage: 6 

VDC

Operating Water
Pressure: 

45 psi min running; 80 psi max static.

 

Note: Installations with running pressure below 
recommended value may void warranty.

Automatic Flush: 

Flush 24 hours after the last use (switchable)

Maintenance 
Override: Magnetic 

fl ush triggering or inhibiting switch (see 

page 2)

Indicator Light: 

Range Adjustment (fi rst ten minutes)

 

Low Battery Indication (see page 2)

 Refl ective surface in view for greater than 30 minutes 

(see page 2)

Reset Button: 

To restart ten-minute range adjustment cycle

 

To enter option changes in the electronics after 
repositioning switches (see page 2)

Operating
Temperature: 

35°F to 104°F (2°C to 40°C)

Outside Case
and Cover: 

Chrome-Plated Die Casting

Operation

1.  Invisible light rays are continually emitted from the RetroFlush sensor.
2. When the user comes into range of  the sensor’s detection zone (range 

12 to 60 inches), the beam is refl ected back to the sensor’s receiver 
and converted to a low voltage electrical  signal  that activates an 
eight-second time delay circuit. This time delay prevents the unit from 
incidental activation in the restroom. Once this circuit is completed, the 
output circuit is signaled and continues in a “hold” mode for as long as 
the user remains within range of the sensor. User must be in sensor 
range for at least 10 seconds.

3. After the user moves away from the sensor, the absence of refl ected light 

starts an electrical “one-time” signal that operates the solenoid operator 
(6 VDC) and activates the fl ushing cycle to fl ush the fi xture. The circuit 
then automatically resets for the next user.

IMPORTANT NOTE

: The Zurn AquaVantage is engineered for quiet 

operation. Excessive water fl ow creates noise, while too little water 
fl ow may not satisfy the needs of the fi xture. Proper adjustment of stop 
valve is made when:

1. The plumbing fi xture is cleansed after each fl ush without splashing water 

out from the lip.

2. A quiet fl ushing cycle is achieved. After adjustment of the stop adjusting 

screw: replace the Zurn stop cap screw cover.

Care and Cleaning Instructions

DO NOT

 use abrasive or chemical cleaners to clean fl ush valve as they 

will dull the luster and attack the chrome or special decorative fi nishes. 
Use only mild soap and water, then wipe dry with a clean cloth or towel. 
While cleaning the bathroom tile and fl oor,  the fl ush valve and electronics 
should be protected from splattering of water, cleaner, acids and cleaning 
fl uids that can damage the sensor fl ush valve.  

Problem

Cause

Solution

Valve will not operate

Water supply shut off

Turn on water supply, check control stop and supply

Sensor range adjustment set too short

Increase sensor range adjustment

Expired batteries

Replace batteries

Diffi culty with electronic control module

Check batteries.  If problem continues, consult factory.

Solenoid problem

Replace solenoid

Red light fl ashes when 
user is not present

Obstruction in front of the sensor

Remove the obstruction

Sensor range set too long
Sensor is detecting opposite wall

Shorten sensor range

Red light fl ashes when 
user comes into view

Batteries about to expire

Replace batteries

Unit is in start-up mode

Light will stop in 10 minutes

Range adjustment reset button was 
pressed.

Light will stop in 10 minutes

Valve does not shut off

Dirt in diaphragm bypass hole

Clean bypass hole

Dirt or debris at valve seat or sealing area

Clean valve seat and clean diaphragm kit

Riser tube fi lter plugged

Clean fi ltered riser tub

Solenoid problem

Replace solenoid

Not enough water to 
fi xture

Wrong RetroFlush unit installed

Check fi xture requirements.  Install proper RetroFlush 
unit.

Control stop improperly adjusted 

Adjust control stop

Diaphragm damaged

Replace diaphragm

Low running pressure

Adjust water supply to ensure 45 psi running pressure 
at each valve.  Move switch #1 to the off position and 
press the RESET button for more water per fl ush.

Solenoid problem

Replace solenoid

Too much water for 
fi xture

Wrong RetroFlush unit installed

Check fi xture requirements.  Install proper RetroFlush 
unit.

Control stop improperly adjusted 

Adjust control stop

Dirt in diaphragm bypass hole

Clean bypass hole

Customer wishes to realize maximum water 
savings

Change out yellow volume control to red volume 
control included with kit.

Water drips from 
chrome cap

The plastic cover has been cracked by 
freezing or abuse

Replace plastic cover

The chrome cap is not tight

Tighten chrome cap

TROUBLESHOOTING GUIDE

ZURN INDUSTRIES, LLC 

5900 Elwin Buchanan Drive, Sanford, NC U.S.A. 27330, Ph. 1-800-997-3876, Fax 919-775-3541

In Canada: 

ZURN INDUSTRIES LIMITED 

7900 Goreway Drive, Unit 10, Brampton, Ontario L6T 5W6, Ph. 905-405-8272, Fax 905-405-1292

Rev. D 

Date: 11/14/2018 

C.N. No. SAN008311 

Prod. | Dwg. No. FV62

Summary of Contents for AquaSense AV ZR Series

Page 1: ...provided we are notified in writing of such defect and the goods are returned to us prepaid at Sanford NC with evidence that they have been properly maintained and used in accordance with instructions We shall not be responsible for any labor charges or any loss injury or damages whatsoever including incidental or consequential damages The sole and exclusive remedy shall be limited to the replacem...

Page 2: ...nut with a wrench Do not turn water on until batteries are inserted see 8 2 Prior to inserting the flush valve tailpiece into stop valve be certain that the o ring seal is located in o ring seal groove at the end of the tailpiece and the locking nut and locking snap ring are located as shown Care should be taken not to damage the o ring when inserting the tailpiece into the stop valve If lubricati...

Page 3: ...he final setting for urinals should be such that the fixture will not overflow when the valve is actuated twenty times in succession The stop cap screw cover should be replaced after final adjustments have been made Sensor Range Adjustment The range is factory set for urinal or closet applications If necessary the sensor range can be adjusted Pressing the reset button will initiate a ten minute se...

Page 4: ...ent For the first ten minutes of operation the red light will flash when an object is detected 2 Low Battery Indication When the batteries have reached the caution level the red light will flash when an object is detected 3 Abnormal Reflection Indicator When a reflective surface is present for longer than 30 minutes the red light will flash indicating that an object is being continuously detected ...

Page 5: ...age while reassembling the unit Insert the solenoid Item 7 into the housing and install the solenoid retaining clip Item 8 When installed properly the end of the solenoid housing is free to rotate Note Be careful not to crimp the battery wires while inserting the solenoid into the housing 3 Reinstall the RetroFlush sensor assembly Item 9 onto the flush valve using a wrench Be sure the solenoid inl...

Page 6: ...Item 8 When installed properly the end of the solenoid is free to rotate Notes Be careful not to crimp the battery wires while inserting the solenoid into the housing Also be sure the blue and gray wires of the solenoid valve extend into the sensor module cavity of the housing Item 9 6 Connect the sensor module blue and gray wires to the solenoid valve blue and gray wires and reinstall the sensor ...

Page 7: ...lume Control 3 5 GpF 3 PR6000 EU25 WS Low Consumption Urinal Volume Control 1 6 GpF 3 PR6000 EU25 WS1 Diaphragm PR6000 E15A Sensor Module Replacement Kits Product No Solenoid Valve Replacement Kit 8 9 10 PR6000 M Sensor Lens Kit 12 PR6000 SC High Volume Sensor 13 PR6000 SH Low Volume Sensor 13 PR6000 SL Battery Holder with Connecting Cable 15 16 PR6000 MB Metal Housing 11 PR6000 MH Vandal Resistan...

Page 8: ...e or chemical cleaners to clean flush valve as they will dull the luster and attack the chrome or special decorative finishes Use only mild soap and water then wipe dry with a clean cloth or towel While cleaning the bathroom tile and floor the flush valve and electronics should be protected from splattering of water cleaner acids and cleaning fluids that can damage the sensor flush valve Problem C...

Reviews: