1
1. Purpose and Features
1.1 Purpose
The surgical microscope OMS2360, OMS2365 (OMS2360, OMS2365) is applicable to the
microsurgery and fine examination of dental clinical subjects (except ophthalmology).
1.2 main structure composition
OMS2360, OMS2365 composed of the optical system, including the variable
magnification zoom lens, the binocular eyepiece, as well as the lighting system, stents,
electrical devices, can all be configured according to different application requirements
along with the corresponding accessories.
1.3 features
OMS2360, OMS2365 can enable teaching through built in photography and video capture
accessories.
OMS2360, OMS2365 use 190 ° adjustable binoculars eyepieces that can be angled to match
the desired positioning for optimal operator viewing comfort.
A specially designed LED light source provides the maximum intensity of illumination needed
to work confidently.
OMS2360 for six times system, OMS2365 to five times.The eyepiece is wide Angle design, the
optical system image is clear, the field is big, the field is long.The built-in zoom lens provides a
larger range of work distance adjustment for doctors.
Built-in color filters with a yellow filter and green filter, Yellow for anti polymerization, green for
surgery.
Stepless brightness adjustment knobs conveniently located on the instrument allow for the
exact level of illumination required to be easily selected.
The cross arm is designed with a spring balance, so that the microscope stays at the right
operating height after positioning without moving up/down. While the microscope is in regular
use, the LED illumination will remain ON, and if the microscope is moved out of operating
range, the LED illumination will automatically turn off allowing for efficient storage when not in
use.
1.4 performance
After taking magnification into account, the viewing point of reference offset is not greater
than 10% of the diameter of the field.
The uniform illumination of the field of view has a color rendering index greater than 85%,
allowing for more accurate visualization of structures and media.
working noise of 65 dB (A) or less.
The large zoom objective lens working distance error is not more than 15%, for example the
shortest working distance is 200 +/- 30 mm, the longest working distance is 450 + / - 67.5 mm.
1.5 contraindications
No clear contraindications have been found.