5. Maintenance 

5.1 Replacing the consumable parts 

5.1.1Replacing the LED 

Please contact the after-sales service department.                         

5.1.2 Replacing the fuse 

The fuse is integrated with main power input socket. 
Please replace the fuse according to the following steps: 
● Turn off the main power switch. 
● Pull out the power plug from the main power plug [22]. 
● Find the fuse socket on the right side of power plug [22], take out the fuse socket from 

the side with a small screwdriver. 

● Take out the melted fuse. 
● Insert a new fuse and insert the fuse socket back again. 
● Plug in power cable. 
● Turn on the main power switch. 
Specification of fuse

T1.25 A/H250 V for 220VAC 

T2.5 A/H250 V for 110VAC 

 

 

5.1.3 Replacing the power supply cords 

Replacement  the  power  supply  cords  should  by  the  manufacturer  specifies  service 
person. 

 

 

 

 

5.2 Cleaning and Sterilization 

 

 

5.2.1 Cleaning the surface of equipment 

The outer surface of the equipment may be cleaned with a wet cloth. The remaining 

stains can be cleaned off with a mixture of 50% C2H5OH and 50% distilled water. Do not 
use any corrosive detergent or the surface may be damaged. 

5.2.2 Cleaning the surface of the optical lens 

To prevent the dust from staining the lens, never expose the optical lens to air without 

the  objective,  eyepiece  tube  and  eyepiece.  Please  use  the  dust  cover  after  using  the 
microscope. 

Cleaning the surface of the optical lens: To clean the dirt on lens, such like bloodstain, 

Caution

Please use the bulb specially designed for this equipment.

 

Caution:    Dirt on the lens must be cleaned immediately after the surgery. It will be 
hard to clean when the dirt air-dried.

 

Warning: To avoid the unacceptable risk, please use the special cords for 

the equipment.

 

Summary of Contents for OMS2360

Page 1: ...with the European medical device regulation EU 2017 745 Zumax Medical Co Ltd 5 Zhiying Street Suzhou New District Suzhou Jiangsu China 215129 Phone 86 512 66650502 Fax 86 512 66909655 E mail Sales zu...

Page 2: ...ical values of any connecting terminals 4 Only use fuse voltage outputs according to the specifications and rated values stipulated by our product 5 Use the power cable supplied with this instrument o...

Page 3: ...Ground Terminal Electrostatic discharge warning 12 Kg Weight limit Brightness adjustment Video plug Power plug Balance arm regulation Caution Operating instructions Warning...

Page 4: ...10 3 9 Video output connection 10 4 Use of the microscope 11 14 4 1 Necessary working condition 11 4 2 Precautions 11 4 3 Installation and adjustment before Use 11 4 3 1 Adjusting the balance of seco...

Page 5: ...es a larger range of work distance adjustment for doctors Built in color filters with a yellow filter and green filter Yellow for anti polymerization green for surgery Stepless brightness adjustment k...

Page 6: ...2 2 Main components Fig 1 general assembly view 1 3 4 2 6 7 8 9 13 A 5 11 14 10 12...

Page 7: ...3 Fig 3 Right back side view Fig 2 1 A direction view 23 24 25 26 28 29 Fig 2 2 base view 15 16 17 18 19 22 21 20 27...

Page 8: ...4 Fig 5 Assembly the camera Fig 4 Left back side view 32 31 36 34 37 35 38 33...

Page 9: ...5 Fig 6 accessories parts 39 41 40 42 44 43 45...

Page 10: ...Wheel Lock the brakes to fix the microscope in place 12 The second arm plate 13 12 5X eyepiece The eyepiece has a dioptric adjustment ring and a goggles height adjustment ring 14 Focus knob Use this k...

Page 11: ...photo Long kick button to record video kick it again to exit video recording 28 Built in camera joint 29 LED power cable and the joint for power cable 31 Balance arm 32 Illumination brightness adjusti...

Page 12: ...ed in one package Please open the package and take out all parts and assemble them according to the following procedures Fig 7 Assembly drawing of microscope connecting arm The first arm column Base S...

Page 13: ...ease pay attention the star handler fixation screw 3 must be tightened in advance insert the hole to the axis then use jump ring pliers to fix the two axis with jump ring in the circle groove put on t...

Page 14: ...7 Assembly of the LED power cable The LED power cable is fixed on the first and second arm Plug LED power cable into the joint 28 3 8 How to connect the power cable Take out the power cable and plug...

Page 15: ...objective The terminals of LED power cable must be plugged into the right holes of the joint Please do not cover the heat elimination grooves of power source Please pay special attention to the cauti...

Page 16: ...ethod to adjust the second eyepiece Assemble the eyepiece and eyepiece tube back to microscope body tighten the fixation screw 10 Adjust the eye cover until the whole visual field can be observed Adju...

Page 17: ...knob etc have been equipped the sterile covers Microscope head may be equipped with a single use sterile cover according to doctor s requirement 4 5 Using the microscope Please make sure the above ste...

Page 18: ...k near the column Fasten every star knob firmly to fix the arm and the microscope Release the brakes of the wheels While moving the equipment hold the movement handles and make the equipment move slow...

Page 19: ...ds should by the manufacturer specifies service person 5 2 Cleaning and Sterilization 5 2 1 Cleaning the surface of equipment The outer surface of the equipment may be cleaned with a wet cloth The rem...

Page 20: ...2 3 Sterilization All sterilization caps should be sterilized by autoclave The following temperature and time are recommended Sterilization temperature 134 for 10 minutes 5 3 disposal of wastes The w...

Page 21: ...e 800 0139 Rubber Eye Guard 2 pcs 800 0151 New 120 deg coupling 800 0166 Canon Digital Camera Adapter 800 0167 Sony Digital Camera Adapter 800 0168 Nikon Digital Camera Adapter 800 0169 Nikon 1J5 Digi...

Page 22: ...wireless Foot pedal for Micro2D 800 0472 wireless phoht video set for integrated full hd video camera 800 0475 Wireless router Networking cables 800 3010 Nissin MF18 Macro Flash for Canon Nikon SLR ca...

Page 23: ...not inserted to the joint correctly Insert it to the joint correctly Microscope not in working area but in parking position Move the second arm to working area The LED has burnt Contact the after sale...

Page 24: ...ng base column first arm and second arm Second Arm Length 720mm Rotational angle 150 Distance 300mm First Arm Length 500mm Rotational angle 350 Height 1750mm Base dimension 610 600mm Electrical specif...

Page 25: ...9 8 6 9 5 4 9 0 6 4 4 9 2 16 1 11 4 8 8 14 7 10 4 8 1 3 24 2 17 1 13 3 22 1 15 6 12 1 numerical aperture NA 0 3 0 008 0 006 0 005 0 008 0 005 0 004 0 5 0 013 0 009 0 007 0 012 0 008 0 006 0 8 0 017 0...

Page 26: ...6 2 8 4 7 3 3 2 6 1 8 0 5 7 4 4 7 3 5 2 4 0 1 6 12 8 9 0 7 0 11 7 8 3 6 4 2 5 20 1 14 2 11 0 18 3 13 0 10 1 numerical aperture NA 0 4 0 010 0 007 0 005 0 010 0 007 0 005 0 6 0 014 0 010 0 008 0 014 0...

Page 27: ...than those specified with the exception of transducers and cables sold by the MANUFACTURER of the microscope as replacement parts for internal components may result in increased EMISSIONS or decrease...

Page 28: ...61000 4 5 1 kV line s to line s 2 kV line s to earth 1 kV line s to line s 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital environment Voltage dips short in...

Page 29: ...ionary RF transmitters as determined by a site surveya should be less than the compliance level in all frequency ranges b Interference may occur in the vicinity of equipment marked with the following...

Reviews: