ZOLL Propaq MD Battery Pack Manual Download Page 9

9650-0809-01 Rev. A

 Propaq ® MD Battery Pack

 Guide

7

Understanding Battery Faults

The 

Propaq MD Battery Pack 

can detect one or more simultaneously occurring faults that 

render the battery unfit for use.

To determine the faults that affect a battery, follow this procedure when the fault light is on:

The following table lists battery temperature faults and actions you can take to clear these fault 
conditions. All other fault codes require you to remove the battery from service. For more 
information on these fault codes, see the 

Propaq MD Service Manual

Press and hold down the 

Display 

button for 

at least

 

15 seconds.

If the battery does not detect any battery faults, all of the 
LEDs will light for 10 seconds.

The 

Fault 

LED will blink to indicate the code that identifies 

the battery fault (there will be a brief delay before the LED 
blinks).

The fault codes number from 1 to 29. To indicate fault code 
1, the 

Fault 

LED blinks once. To indicate fault code 2, the 

LED blinks twice, and so on.

If the battery detects multiple battery faults, it displays the 
first fault code, pauses briefly, and then displays the next 
fault code, and so on.

Fault ID

Fault Type

Recommended Action

13

Battery Over Temperature 
During Charge.

The battery is at or above its 
maximum temperature for 
charging (140°F/60°C).

Move the battery to a cooler room or out of 
direct sunlight, if possible, and allow the 
battery to cool before continuing to charge the 
battery. 

If the ambient temperature was low when the 
battery reached or exceeded its maximum 
temperature, do not use this battery. 

14

Battery Under Temperature 
during Charge.

The battery is at or below the 
minimum temperature for 
charging (32°F/0°C).

Move the battery to a warmer room (above 
32°F / 0°C) and allow the battery to warm up 
before charging again.

Summary of Contents for Propaq MD Battery Pack

Page 1: ...Propaq MD Battery Pack Guide ITHIUM ON ATTERY 2 9650 0809 01 Rev A...

Page 2: ...al Corporation to determine if additional product information updates are available ZOLL is a registered trademark and SurePower is a trademark of ZOLL Medical Corporation Propaq is a registered trade...

Page 3: ...tery Pack be dropped inspect and test the unit prior to further use following the procedures described in the Maintenance section of this Guide Avoid prolonged exposure of the Propaq MD Battery Pack t...

Page 4: ...fibrillator and battery Cautions Follow all recommended maintenance instructions If a problem occurs obtain service immediately Do not use alcohol or ketones MEK acetone etc to clean Propaq MD Battery...

Page 5: ...ttery s calibration status Battery Fault Indications Figure 1 1 ZOLL Propaq MD Battery Pack Note ZOLL ships the Propaq MD Battery Pack in a partially charged state and you must fully charge the batter...

Page 6: ...mately 10 seconds to indicate the battery s estimated run time If the battery is fully charged the LED over each of the five runtime indicator icons illuminates As the battery discharges the number of...

Page 7: ...ndications This is the Recalibration icon This LED lights for approximately 10 seconds after you press and release the Display button if the battery needs to be calibrated If the Recalibration LED lig...

Page 8: ...untime indicator displays the battery runtime When you release the Display button the runtime indicator displays the battery capacity for 2 seconds then redisplays the battery runtime During the 2 sec...

Page 9: ...ault there will be a brief delay before the LED blinks The fault codes number from 1 to 29 To indicate fault code 1 the Fault LED blinks once To indicate fault code 2 the LED blinks twice and so on If...

Page 10: ...t in 17 8 Length x 9 3 Width x 3 Height cm Weight 1 4 lb 64 kg Voltage 11 1 V Capacity 6 6 Ah when charged to 12 6V Operating Temperature 32 F to 122 F 0 C to 50 C Optimal Charging Temperature 59 F to...

Reviews: