ZOLL Propaq MD Battery Pack Manual Download Page 7

9650-0809-01 Rev. A

 Propaq ® MD Battery Pack

 Guide

5

Cleaning

Clean the 

Propaq MD Battery Pack

 with a soft damp cloth and mild soap. 

Do not immerse or set the 

Propaq MD Battery Pack 

in fluids of any kind. 

Do not

 use alcohol or 

ketones to clean the 

Propaq MD Battery Pack

.

Maintenance -- Testing and Inspection

Perform a battery check, which verifies that the battery is fully charged and fault free, as part of 
any regularly scheduled unit test. Press the battery’s 

Display

 button to perform the battery 

check.

As part of your defibrillator/monitor maintenance program, verify that the battery’s LEDs work 
(press the 

Display

 button to test LEDs) -- if the LEDs do not work, remove the battery from 

service. Also check the battery casing for cracks or other damage -- if the battery is cracked or 
damaged, remove it from service.

Propaq MD Battery Pack

 

Icon

LED Indications

This is the 

Recalibration

 icon. 

This LED lights for approximately 10 seconds (after you 
press and release the 

Display

 button) if the battery needs to 

be calibrated. If the 

Recalibration

 LED lights, the runtime 

indicator will 

not

 display run time for that battery. For best 

performance of the battery, you should recalibrate the 
battery as soon as possible.

To manually recalibrate the 

Propaq MD Battery Pack

, you 

can insert the battery into the 

SurePower Charger Station 

(using the 

Propaq Battery Adaptor

) and perform a 

ManualTest

 (for more information, see the 

ZOLL 

SurePower Charger Station 

Operator’s Guide

). After you 

recalibrate the battery, the 

Recalibration 

LED will only flash 

momentarily when you press the 

Display

 button.

This is the 

Fault

 Icon. 

If the 

Propaq MD Battery Pack

 detects a battery fault, this 

LED lights for approximately 10 seconds after you press the 

Display

 button.

If the 

Fault

 LED lights, the runtime indicator will 

not

 display 

run time for that battery.

See the following section for more information on battery 
faults.

Summary of Contents for Propaq MD Battery Pack

Page 1: ...Propaq MD Battery Pack Guide ITHIUM ON ATTERY 2 9650 0809 01 Rev A...

Page 2: ...al Corporation to determine if additional product information updates are available ZOLL is a registered trademark and SurePower is a trademark of ZOLL Medical Corporation Propaq is a registered trade...

Page 3: ...tery Pack be dropped inspect and test the unit prior to further use following the procedures described in the Maintenance section of this Guide Avoid prolonged exposure of the Propaq MD Battery Pack t...

Page 4: ...fibrillator and battery Cautions Follow all recommended maintenance instructions If a problem occurs obtain service immediately Do not use alcohol or ketones MEK acetone etc to clean Propaq MD Battery...

Page 5: ...ttery s calibration status Battery Fault Indications Figure 1 1 ZOLL Propaq MD Battery Pack Note ZOLL ships the Propaq MD Battery Pack in a partially charged state and you must fully charge the batter...

Page 6: ...mately 10 seconds to indicate the battery s estimated run time If the battery is fully charged the LED over each of the five runtime indicator icons illuminates As the battery discharges the number of...

Page 7: ...ndications This is the Recalibration icon This LED lights for approximately 10 seconds after you press and release the Display button if the battery needs to be calibrated If the Recalibration LED lig...

Page 8: ...untime indicator displays the battery runtime When you release the Display button the runtime indicator displays the battery capacity for 2 seconds then redisplays the battery runtime During the 2 sec...

Page 9: ...ault there will be a brief delay before the LED blinks The fault codes number from 1 to 29 To indicate fault code 1 the Fault LED blinks once To indicate fault code 2 the LED blinks twice and so on If...

Page 10: ...t in 17 8 Length x 9 3 Width x 3 Height cm Weight 1 4 lb 64 kg Voltage 11 1 V Capacity 6 6 Ah when charged to 12 6V Operating Temperature 32 F to 122 F 0 C to 50 C Optimal Charging Temperature 59 F to...

Reviews: