ZOLL Propaq MD Battery Pack Manual Download Page 4

I

NTRODUCTION

2

www.zoll.com

9650-0809-01 Rev. A

Do not place foreign objects into battery sockets.

Do not use the 

Propaq MD Battery Pack

 in the presence of flammable agents (such as 

gasoline), oxygen-rich atmospheres, or flammable anesthetics. Using the device in the 

presence of flammable agents can cause an explosion. 

Do not immerse or set the 

Propaq MD Battery Pack

 in liquid. Using the battery near or 

within puddles of water may present a shock hazard to the operator, patient, and nearby 

personnel.

Use only ZOLL chargers to charge ZOLL batteries.

If the display indicators on the 

Propaq MD Battery Pack

 are not consistent with the 

indicators on the 

SurePower Charger Station

, a fault condition may be present. Remove the 

battery from service as soon as practical, and verify the correct operation of the charger, 

defibrillator, and battery.

Cautions

Follow all recommended maintenance instructions. If a problem occurs, obtain service 

immediately. 

Do not use alcohol or ketones (MEK, acetone, etc.) to clean 

Propaq MD Battery Pack

s. 

Do not sterilize or autoclave 

Propaq MD Battery Pack

s.

Avoid storage of fully depleted Propaq Battery Packs.

Do not install batteries into a Propaq MD unit that is not plugged into live AC mains when 

storage may exceed 90 days. Battery damage may occur.

If you are disposing of a 

Propaq MD Battery Pack

, contact your local authorities to 

determine the requirements for the recycling and disposal of electrical equipment and follow 

those requirements.

Notes

Propaq MD Battery Packs

 may feel warm to the touch following a large number of shocks 

delivered at maximum energy or after recharging.

You can get additional information about the 

Propaq MD Battery Pack

 and its use from the 

Propaq MD Operator’s Manual 

(for example, labeling information -- symbols used, Serial 

Number format, Date of Manufacture, and so on).

Summary of Contents for Propaq MD Battery Pack

Page 1: ...Propaq MD Battery Pack Guide ITHIUM ON ATTERY 2 9650 0809 01 Rev A...

Page 2: ...al Corporation to determine if additional product information updates are available ZOLL is a registered trademark and SurePower is a trademark of ZOLL Medical Corporation Propaq is a registered trade...

Page 3: ...tery Pack be dropped inspect and test the unit prior to further use following the procedures described in the Maintenance section of this Guide Avoid prolonged exposure of the Propaq MD Battery Pack t...

Page 4: ...fibrillator and battery Cautions Follow all recommended maintenance instructions If a problem occurs obtain service immediately Do not use alcohol or ketones MEK acetone etc to clean Propaq MD Battery...

Page 5: ...ttery s calibration status Battery Fault Indications Figure 1 1 ZOLL Propaq MD Battery Pack Note ZOLL ships the Propaq MD Battery Pack in a partially charged state and you must fully charge the batter...

Page 6: ...mately 10 seconds to indicate the battery s estimated run time If the battery is fully charged the LED over each of the five runtime indicator icons illuminates As the battery discharges the number of...

Page 7: ...ndications This is the Recalibration icon This LED lights for approximately 10 seconds after you press and release the Display button if the battery needs to be calibrated If the Recalibration LED lig...

Page 8: ...untime indicator displays the battery runtime When you release the Display button the runtime indicator displays the battery capacity for 2 seconds then redisplays the battery runtime During the 2 sec...

Page 9: ...ault there will be a brief delay before the LED blinks The fault codes number from 1 to 29 To indicate fault code 1 the Fault LED blinks once To indicate fault code 2 the LED blinks twice and so on If...

Page 10: ...t in 17 8 Length x 9 3 Width x 3 Height cm Weight 1 4 lb 64 kg Voltage 11 1 V Capacity 6 6 Ah when charged to 12 6V Operating Temperature 32 F to 122 F 0 C to 50 C Optimal Charging Temperature 59 F to...

Reviews: