Cool Line

® 

Intravascular Heat Exchange Catheter (Custom Luer) 

Instructions for Use 

CL-2295A/8700-0781-01 

 

2 of 10

 

ZOLL

 

6.

 

 

Warnings and Precautions 

WARNING: The catheter and Start-Up Kit could 
potentially misconnect with other devices with small bore 
connectors. Such misconnection errors could result in 
patient injury or death. 

CAUTION: The custom Luers on the catheter and SUK 
may reduce the risk of misconnections but still have the 
potential for misconnections with these specific medical 
device applications: Breathing Systems & Driving Gases 
applications, Enteral & Gastric applications, Urethral & 
Urinary applications, Limb Cuff Inflation applications, 
Neuraxial applications, and Intravascular or Hypodermic 
applications.  Always use caution when connecting ZOLL 
catheters and SUK’s to these and other medical device 
applications. 

CAUTION:  Ensure that the catheter and/or Start-Up Kit 
are not connected to an IV or other medical device. 

1.

 

SINGLE USE ONLY. 

The product is designed for single 

use only. Do not resterilize or reuse. Do not reinsert, once 
removed from the patient. Do not alter the catheter in any 
way. Maximum use period: 7 days. 

2.

 

The use of intravascular cooling devices controls fever, 
including fever due to sepsis.  Care must be taken to assess 
patients for sepsis. 

3.

 

Alcohol and acetone can weaken the structure of the shaft 
material. Care should therefore be taken when infusing drugs 
containing alcohol or when using alcohol or acetone when 
performing routine catheter care and maintenance. Alcohol 
should not be used to declot the catheter.  

4.

 

Use of a syringe smaller than 10 ml to irrigate or declot an 
occluded catheter may cause intraluminal leakage or 
catheter rupture.  

5.

 

Caution:  If blood is observed within the sterile saline 
circuit, stop the procedure.

  

6.

 

The catheter is coated with heparin. This may induce or 
aggravate pre-existing heparin-induced thrombocytopenia 
(HIT).  

7.

 

Central venous catheterization should only be performed 
by well-trained personnel well-versed in anatomical 
landmarks and safe technique. Personnel should also have 
knowledge of potential complications. 

8.

 

The catheter should be placed via a jugular, subclavian, or 
femoral vein approach only. 

9.

 

Do not allow the catheter to be placed into the right atrium 
or right ventricle. If placed via the jugular or subclavian 
veins, the catheter should be positioned so that the distal tip 
of the catheter is in the superior vena cava above its junction 
with the right atrium and parallel to the vessel wall. X-ray 
examination should be used to ensure that the catheter is not 
in the right atrium or ventricle. The distal tip of the catheter 
should be positioned at a level above either the azygos vein 
or the carina of the trachea, whichever is better visualized. 

10.

 

If placed via the femoral vein, the catheter should be 
positioned so that its distal tip is in the inferior vena cava, 
below its junction with the right atrium and parallel to the 
vessel wall. 

11.

 

Possible complications with central venous catheters 
include: atrial or ventricular perforation, cardiac 

tamponade, air embolism, catheter embolism, thoracic 
duct laceration, bacteremia, septicemia, thrombosis, 
inadvertent arterial puncture, hematoma formation, 
hemorrhage, nerve damage and dysrhythmias. 

12.

 

All Luer-Lock connections and covers must be securely 
tightened to prevent air embolism or fluid or blood loss. 

13.

 

Never use excessive force in moving the catheter or 
guidewire. If resistance is encountered, an x-ray should 
be performed to identify the reason for the resistance. 

14.

 

Passage of the guidewire into the right heart can cause 
dysrhythmias, right bundle branch block, vessel wall, 
atrial or ventricular perforation. 

15.

 

Use only sterile normal saline for catheter priming. It is 
the circulating fluid in the catheter. 

16.

 

The catheter should be routinely inspected for flow rate, 
security of dressing, correct catheter position and for 
secure Luer-Lock connections. Use the centimeter 
markings to identify if the catheter position has changed. 

17.

 

Only x-ray examination can ensure that the catheter tip 
has not entered the heart or no longer lies parallel to the 
vessel wall. If the catheter position has changed, perform 
an x-ray examination to confirm catheter tip position. 

18.

 

For blood sampling, temporarily shut off remaining 
infusion ports through which solutions are being infused. 

19.

 

Use only a 30 cc or smaller syringe for blood sampling. 

20.

 

Use care when infusing drugs that may be affected by 
cool temperatures (as low as 4ºC). Mannitol-containing 
solutions are temperature sensitive and must not be 
delivered through the catheter except for a rapid push of 
up to a concentration of 20% mannitol solution, followed 
by saline flush. Higher than a 20% concentration of 
mannitol or drip or infusion pump delivery of mannitol 
must be done via a separate line. 

21.

 

Do not infuse into the orange Luer-Lock connections. 

22.

 

Use only the ZOLL suture tab and clip provided in the kit 
to prevent catheter damage. 

23.

 

Not intended for pediatric or neonatal use. 

24.

 

Cardiac Tamponade: Placement of indwelling catheters in 
the right atrium is a practice that may lead to cardiac 
perforation and tamponade. Practitioners placing central 
venous catheters must be aware of this potentially fatal 
complication before advancing the catheter too far relative to 
patient size. The actual position of the tip of the indwelling 
catheter should be confirmed by x-ray after insertion. 
Central venous catheters should not be placed in the right 
atrium unless specifically required for special relatively 
short-term procedures, such as aspiration of air emboli 
during neurosurgery. Such procedures are nevertheless risk 
prone and should be closely monitored and controlled. 

25.

 

When connecting infusion sets/injection systems to 
ZOLL catheters, do not exceed 100 psi/689 kPa. 

WARNING: INTRALUMINAL LEAKAGE

 

Intraluminal 

leakage between the saline Luer and infusion Luers is an 
uncommon but potential catheter failure. In the event of such 
a failure, sterile saline from the cooling circuit will be 
introduced into the patient. Intraluminal leakage will usually 
be associated with a fluid loss alarm, which will stop the 
system. ALWAYS INVESTIGATE FLUID LEVEL 
ALARMS. The cooling circuit is a closed loop system – 
usually fluid loss alarms indicate a breach somewhere in this 
closed loop. With any fluid loss alarm, check both the 
integrity of the catheter and the Start-Up Kit (see below).

 

Summary of Contents for Cool Line CL-2295A/8700-0781-01

Page 1: ... Indications for Use The Cool Line Catheter Model CL 2295A when used with the ZOLL Thermal Regulation System is indicated for use in fever reduction as an adjunct to other antipyretic therapy in adult patients with cerebral infarction and intracerebral hemorrhage who require access to the central venous circulation and who are intubated and sedated Safety and Efficacy Considerations Central venous...

Page 2: ...t laceration bacteremia septicemia thrombosis inadvertent arterial puncture hematoma formation hemorrhage nerve damage and dysrhythmias 12 All Luer Lock connections and covers must be securely tightened to prevent air embolism or fluid or blood loss 13 Never use excessive force in moving the catheter or guidewire If resistance is encountered an x ray should be performed to identify the reason for ...

Page 3: ...Materials Required They are not intended to connect to standard Luer Lock syringes or other standard Luer Lock connectors 3 Place the patient in a slight Trendelenburg position as tolerated to reduce the risk of air embolism If a femoral approach is used place the patient in a supine position 4 Prep and drape the puncture site as required 5 Caution Always prime the catheter before it is inserted i...

Page 4: ...LL suture tab and clip can also be used as an additional attachment point Ensure that the catheter body is secure and does not slide 19 Caution Use only the ZOLL suture tab and clip provided in the kit Catheter damage may result if other tabs or clips are used 20 Caution Do not suture directly to the outside diameter of the catheter to minimize the risk of cutting or damaging the catheter or imped...

Page 5: ... the sutures from the suture site 6 Slowly remove the catheter from patient As the catheter exits the site apply pressure with a dressing impermeable to air e g Vaseline gauze 7 Warning Do not move the catheter if resistance is felt Check to ensure that the IN and OUT Luers of the cooling circuit are NOT capped If they are capped uncap them deflate the balloon and try removing the catheter again I...

Page 6: ...e at all times Dispenser Every guidewire is provided in a dispenser package Remove the guidewire anti migration clip before dispensing the guidewire Remove the guidewire protective cap immediately prior to guidewire use Prepare the guidewire prior to insertion It is recommended that the dispenser be filled with heparinized solutions e g saline or dextrose to bathe the guidewire during insertion Th...

Page 7: ...oximal Infusion Lumen 5 OUT 6 Radiopaque Shaft 7 Depth Markers 8 Radiopaque Marker 9 Cold Saline Return Port 10 Balloon 11 Cold Saline Supply Port 12 Proximal Infusion Lumen Side Port 13 Balloon 14 Cold Saline Supply Port 15 Medial Infusion Lumen Side Port 16 Radiopaque Tapered Soft Tip 17 Distal Guidewire and Infusion Port Cool Line CL 2295 Catheter 18 IN 19 OUT Catheter Cross Section ...

Page 8: ...Cool Line Intravascular Heat Exchange Catheter Custom Luer Instructions for Use CL 2295A 8700 0781 01 8 of 10 ZOLL ...

Page 9: ...Cool Line Intravascular Heat Exchange Catheter Custom Luer Instructions for Use CL 2295A 8700 0781 01 9 of 10 ZOLL ...

Page 10: ...Cool Line Intravascular Heat Exchange Catheter Custom Luer Instructions for Use CL 2295A 8700 0781 01 10 of 10 ZOLL SilvaSorb ...

Reviews: