background image

Cool Line

® 

Intravascular Heat Exchange Catheter (Custom Luer) 

Instructions for Use 

CL-2295A/8700-0781-01 

 

3 of 10

 

ZOLL

 

To check the integrity of the catheter 

1.

 

Stop operation of the Coolgard 3000/ 
Thermogard XP System. 

2.

 

Disconnect the Start-Up Kit from the catheter. Properly 
cap both the catheter and Start-Up Kit using an aseptic 
technique. 

3.

 

Fill a sterile 10 ml slip tip syringe with sterile saline.  

4.

 

Connect the syringe to the IN Luer of the catheter and 
disconnect the OUT cap. Infuse the 10 ml of saline– it 
should flow out the OUT Luer.  

5.

 

Cap the OUT Luer and pull 5 cc of vacuum. Sustain for at 
least 10 seconds. Approximately 4 ml of saline, but not 
blood, should enter the syringe and you should be able to 
maintain the vacuum.  

6.

 

Ease the vacuum and recap the IN Luer.  

To check the integrity of the Start-Up Kit 

1.

 

Look for obvious leakage. 

2.

 

Remove the tubing from the pump raceway and inspect 
for damage (return it to position if not damaged).  

3.

 

Check along the tubing from the pump to the patient for 
sources of fluid loss.  

 

Look for damage to the tubing and/or the presence of 
air within the tubing. 

 

Inspect, and tighten as necessary, each Luer fitting 
(do not use instruments to tighten Luer fittings). 

 

Note: Condensation on the exterior of the tubing is 
normal. 

4.

 

Similarly, check the tubing that returns to the pump from 
the patient. Examine the saline bag to ensure that it has not 
been accidentally compromised (for example, the spike 
may have damaged the bag wall).  

5.

 

Trace the tubing from the saline bag back to the pump.  

More warnings and precautions are located in following 
instructions. 

Materials Required 

Quantity  Description 

Cool Line

 

Kit for percutaneous introduction 

500 cc bag of sterile normal saline 

Coolgard 3000/Thermogard XP Start-Up Kit 
(provided separately) 

 

6 ft (183 cm) Standard Tubing or 

 

9 ft (274 cm) Extended Tubing 

Coolgard 3000/Thermogard XP

 

System 

Catheter Preparation and Insertion 

Note

: The catheter has a radiopaque marker band to assist in 

identification of the catheter during and after insertion when 
viewed using x-ray equipment. The proximal end of the 
proximal balloon has one marker band. The tip of the catheter 
contains barium sulfate to make it radiopaque.  

Use sterile technique. 

1.

 

Caution:  Use jugular, subclavian, or femoral vein 
approach only. 

2.

 

Caution: The IN and OUT Luer-Locks on the catheter 
are custom manufactured and are intended to connect 
only with the ZOLL Start Up Kits listed in Materials 
Required. They are not intended to connect to standard 
Luer-Lock syringes or other standard Luer-Lock 
connectors.  

3.

 

Place the patient in a slight Trendelenburg position as 
tolerated to reduce the risk of air embolism. If a femoral 
approach is used, place the patient in a supine position. 

4.

 

Prep and drape the puncture site as required. 

5.

 

Caution:  Always prime the catheter before it is 
inserted into the patient. 

6.

 

Carefully remove the catheter from the package, leaving 
on the catheter membrane cover. 

Catheter Preparation Procedure 

1.

 

Remove the caps from the IN and OUT Luers. With the 
catheter cover in place, fill a 5 cc or larger syringe with 
sterile saline and attach the syringe to the female IN 
Luer. 

2.

 

Warning: Never inject positive pressure into the IN 
Luer with the OUT Luer cap in place.  

3.

 

Gently inject saline through the catheter until it begins to 
exit from the OUT Luer.  

4.

 

Using a 5 cc or larger syringe, flush the distal and 
proximal infusion Luers with sterile saline. Clamp or 
attach injection caps to the proximal infusion Luer. 
Leave the distal Luer uncapped for guidewire passage.  

5.

 

Remove the catheter membrane cover. If there is 
resistance in removing the membrane cover from the 
catheter, flush the membrane cover with sterile saline. 
Inspect the catheter to ensure that air has been purged 
from the heat exchange membrane. Inspect the catheter 
for leaks.  

6.

 

Caution:  Avoid excessive wiping of the coated 
catheter. Avoid wiping the catheter with dry gauze, as 
this may damage the catheter coating.  Avoid using 
alcohol, antiseptic solutions, or other solvents to pre-
treat the catheter, because this may cause 
unpredictable changes in the coating, which could 
affect the device safety and performance. 

7.

 

Warning: Do not cut the catheter to alter length.

  

Catheter Insertion 

1.

 

Obtain jugular, subclavian, or femoral venous access 
using standard percutaneous techniques. Access should 
be maintained with a .032" (0.81 mm) guidewire. See 
Special Instructions for Guidewires

2.

 

Warning: Do not attempt to re-insert a partially or 
completely withdrawn OTN (over the needle) 
introducer needle from its catheter. 

3.

 

Caution: Do not use a guidewire larger than .032" 
(0.81 mm) with the Cool Line catheter. 

4.

 

Holding the guidewire in place, remove the introducer 
catheter. 

Caution: Maintain a firm grip on the 

guidewire at all times.  

5.

 

Enlarge the cutaneous puncture site with the cutting edge 
of the scalpel positioned away from the guidewire. 

Warning: Do not cut the guidewire.

 Use a vessel 

dilator to enlarge the site as required. Do not leave the 
vessel dilator in place as an indwelling catheter to 
minimize the risk of vessel wall perforation 

 

 

Summary of Contents for Cool Line CL-2295A/8700-0781-01

Page 1: ... Indications for Use The Cool Line Catheter Model CL 2295A when used with the ZOLL Thermal Regulation System is indicated for use in fever reduction as an adjunct to other antipyretic therapy in adult patients with cerebral infarction and intracerebral hemorrhage who require access to the central venous circulation and who are intubated and sedated Safety and Efficacy Considerations Central venous...

Page 2: ...t laceration bacteremia septicemia thrombosis inadvertent arterial puncture hematoma formation hemorrhage nerve damage and dysrhythmias 12 All Luer Lock connections and covers must be securely tightened to prevent air embolism or fluid or blood loss 13 Never use excessive force in moving the catheter or guidewire If resistance is encountered an x ray should be performed to identify the reason for ...

Page 3: ...Materials Required They are not intended to connect to standard Luer Lock syringes or other standard Luer Lock connectors 3 Place the patient in a slight Trendelenburg position as tolerated to reduce the risk of air embolism If a femoral approach is used place the patient in a supine position 4 Prep and drape the puncture site as required 5 Caution Always prime the catheter before it is inserted i...

Page 4: ...LL suture tab and clip can also be used as an additional attachment point Ensure that the catheter body is secure and does not slide 19 Caution Use only the ZOLL suture tab and clip provided in the kit Catheter damage may result if other tabs or clips are used 20 Caution Do not suture directly to the outside diameter of the catheter to minimize the risk of cutting or damaging the catheter or imped...

Page 5: ... the sutures from the suture site 6 Slowly remove the catheter from patient As the catheter exits the site apply pressure with a dressing impermeable to air e g Vaseline gauze 7 Warning Do not move the catheter if resistance is felt Check to ensure that the IN and OUT Luers of the cooling circuit are NOT capped If they are capped uncap them deflate the balloon and try removing the catheter again I...

Page 6: ...e at all times Dispenser Every guidewire is provided in a dispenser package Remove the guidewire anti migration clip before dispensing the guidewire Remove the guidewire protective cap immediately prior to guidewire use Prepare the guidewire prior to insertion It is recommended that the dispenser be filled with heparinized solutions e g saline or dextrose to bathe the guidewire during insertion Th...

Page 7: ...oximal Infusion Lumen 5 OUT 6 Radiopaque Shaft 7 Depth Markers 8 Radiopaque Marker 9 Cold Saline Return Port 10 Balloon 11 Cold Saline Supply Port 12 Proximal Infusion Lumen Side Port 13 Balloon 14 Cold Saline Supply Port 15 Medial Infusion Lumen Side Port 16 Radiopaque Tapered Soft Tip 17 Distal Guidewire and Infusion Port Cool Line CL 2295 Catheter 18 IN 19 OUT Catheter Cross Section ...

Page 8: ...Cool Line Intravascular Heat Exchange Catheter Custom Luer Instructions for Use CL 2295A 8700 0781 01 8 of 10 ZOLL ...

Page 9: ...Cool Line Intravascular Heat Exchange Catheter Custom Luer Instructions for Use CL 2295A 8700 0781 01 9 of 10 ZOLL ...

Page 10: ...Cool Line Intravascular Heat Exchange Catheter Custom Luer Instructions for Use CL 2295A 8700 0781 01 10 of 10 ZOLL SilvaSorb ...

Reviews: