background image

Cool Line

® 

Intravascular Heat Exchange Catheter (Custom Luer) 

Instructions for Use 

CL-2295A/8700-0781-01 

 

10 of 10

 

ZOLL

 

 

 

 

 

 

SilvaSorb 

 

Summary of Contents for Cool Line CL-2295A/8700-0781-01

Page 1: ... Indications for Use The Cool Line Catheter Model CL 2295A when used with the ZOLL Thermal Regulation System is indicated for use in fever reduction as an adjunct to other antipyretic therapy in adult patients with cerebral infarction and intracerebral hemorrhage who require access to the central venous circulation and who are intubated and sedated Safety and Efficacy Considerations Central venous...

Page 2: ...t laceration bacteremia septicemia thrombosis inadvertent arterial puncture hematoma formation hemorrhage nerve damage and dysrhythmias 12 All Luer Lock connections and covers must be securely tightened to prevent air embolism or fluid or blood loss 13 Never use excessive force in moving the catheter or guidewire If resistance is encountered an x ray should be performed to identify the reason for ...

Page 3: ...Materials Required They are not intended to connect to standard Luer Lock syringes or other standard Luer Lock connectors 3 Place the patient in a slight Trendelenburg position as tolerated to reduce the risk of air embolism If a femoral approach is used place the patient in a supine position 4 Prep and drape the puncture site as required 5 Caution Always prime the catheter before it is inserted i...

Page 4: ...LL suture tab and clip can also be used as an additional attachment point Ensure that the catheter body is secure and does not slide 19 Caution Use only the ZOLL suture tab and clip provided in the kit Catheter damage may result if other tabs or clips are used 20 Caution Do not suture directly to the outside diameter of the catheter to minimize the risk of cutting or damaging the catheter or imped...

Page 5: ... the sutures from the suture site 6 Slowly remove the catheter from patient As the catheter exits the site apply pressure with a dressing impermeable to air e g Vaseline gauze 7 Warning Do not move the catheter if resistance is felt Check to ensure that the IN and OUT Luers of the cooling circuit are NOT capped If they are capped uncap them deflate the balloon and try removing the catheter again I...

Page 6: ...e at all times Dispenser Every guidewire is provided in a dispenser package Remove the guidewire anti migration clip before dispensing the guidewire Remove the guidewire protective cap immediately prior to guidewire use Prepare the guidewire prior to insertion It is recommended that the dispenser be filled with heparinized solutions e g saline or dextrose to bathe the guidewire during insertion Th...

Page 7: ...oximal Infusion Lumen 5 OUT 6 Radiopaque Shaft 7 Depth Markers 8 Radiopaque Marker 9 Cold Saline Return Port 10 Balloon 11 Cold Saline Supply Port 12 Proximal Infusion Lumen Side Port 13 Balloon 14 Cold Saline Supply Port 15 Medial Infusion Lumen Side Port 16 Radiopaque Tapered Soft Tip 17 Distal Guidewire and Infusion Port Cool Line CL 2295 Catheter 18 IN 19 OUT Catheter Cross Section ...

Page 8: ...Cool Line Intravascular Heat Exchange Catheter Custom Luer Instructions for Use CL 2295A 8700 0781 01 8 of 10 ZOLL ...

Page 9: ...Cool Line Intravascular Heat Exchange Catheter Custom Luer Instructions for Use CL 2295A 8700 0781 01 9 of 10 ZOLL ...

Page 10: ...Cool Line Intravascular Heat Exchange Catheter Custom Luer Instructions for Use CL 2295A 8700 0781 01 10 of 10 ZOLL SilvaSorb ...

Reviews: