CAUTIONS

CAUTION: Federal (U.S.A.) law restricts this device 
to sale, or use by or on the order of a licensed 
medical practitioner. 

Use only ZOLL/Respironics Novametrix 
CAPNOSTAT 3 sensors and LoFlo modules, airway 
adapters, nasal and nasal/oral cannulae with the 
M Series EtCO

2

 option.

The device is protected against interference from 
radio frequency emissions typical of two-way radios 
and cellular phones (digital and analog) used in 
emergency service/public safety activities. Users 
should assess the device’s performance in their 
typical environment of use for the possibility of radio 
frequency interference from high-power sources. 
Radio Frequency Interference (RFI) may be 
observed as shifts in monitor baseline, trace 
compression, display brightness changes or transient 
spikes on the display.

Do NOT sterilize or immerse the CAPNOSTAT 3 
sensor or LoFlo module.

Do NOT reuse or sterilize the disposable airway 
adapter, airway adapter with mouthpiece, 
Capno

2

mask, nasal or nasal/oral sampling 

cannulae, or airway adapters, as system 
performance will be compromised. These items are 
intended for single patient use only. 

Do NOT use a damaged sensor or airway adapter.

Do NOT use the device if it fails to operate properly. 

Do NOT place the mainstream or sidestream airway 
adapters between the ET tube and the breathing 
circuit elbow, as this may allow patient secretions to 
accumulate in the adapter.

Position mainstream airway adapters with windows 
in a vertical, NOT a horizontal, position. This helps 
keep patient secretions from pooling on the 
windows.

Do NOT insert any object other than the sample cell 
into the sample cell receptacle on the LoFlo module.

ZOLL Medical Corporation recommends that the 
CAPNOSTAT 3 airway adapter be removed from the 
circuit whenever aerosolized medication is delivered. 
The increased viscosity of the medications may 
contaminate the adapter windows, requiring 
premature cleaning or replacement of the adapter.

In order to eliminate the potential buildup of CO

inside the storage bag, ensure that the LoFlo module 
exhaust tube vents gasses away from the module 
environment.

To avoid injury to the patient, remove the nasal/oral 
cannula from the patient before cutting the oral 
cannula tip. 

Do NOT apply tension to the sensor or module 
cable.

Periodically inspect the sampling tubing for kinks.

Monitor the capnogram for an elevated baseline. If 
an elevated baseline is observed, verify patient 
condition first. If the care giver determines that the 
patient condition is not contributing to the elevated 
baseline, follow the instructions for zeroing the 
sensor or module detailed in this manual. 

Do NOT store sensors, modules, airway adapters, or 
cannulae at temperatures less than -10° C or greater 
than 55° C.

Do not operate sensors or modules at temperatures 
less than 10° C or greater than 40° C.

Refer servicing to qualified personnel.

Do not use the LoFlo module with M Series units that 
have a software version lower than 38.25, or with 
M Series CCT units that have a software version 
lower than 57.50.

Make sure to insert the protective cap into the LoFlo 
module when it is not in use.

Summary of Contents for Capnostat 3

Page 1: ...9650 0223 01 Rev D TM End Tidal Carbon Dioxide EtCO2...

Page 2: ...l Corporation CAPNOSTAT is a registered trademark and LoFlo is a trademark of Respironics Novametrix LLC Cidex is a registered trademark of Advanced Sterilization Products a Johnson and Johnson Compan...

Page 3: ...mum filtration of fluids and contaminants and protecting the system from aspiration of these fluids In both systems the CO2 sensor generates infrared light and beams it through the airway adapter or s...

Page 4: ...es EtCO2 option Activate oxygen compensation if O2 levels in excess of 60 are introduced Activate N2O compensation if nitrous oxide is introduced into the airway circuit The presence of Desflurane bey...

Page 5: ...h windows in a vertical NOT a horizontal position This helps keep patient secretions from pooling on the windows Do NOT insert any object other than the sample cell into the sample cell receptacle on...

Page 6: ...t and property even if ZOLL has been advised of the possibility of such damages THE WARRANTY SET FORTH HEREIN IS EXCLUSIVE AND ZOLL MEDICAL CORPORATION EXPRESSLY DISCLAIMS ALL OTHER WARRANTIES WHETHER...

Page 7: ...EtCO2 setup as follows Attaching the CAPNOSTAT 3 sensor cable Selecting a mainstream airway adapter Attaching the airway adapter to the CAPNOSTAT 3 sensor Zeroing the mainstream sensor airway adapter...

Page 8: ...ximately 15 seconds The unit displays the message ZERO DONE upon completion of the zeroing Note Do not attempt zeroing for 20 seconds after removing the adapter from the patient s airway This time all...

Page 9: ...ow steps a through c else go to step 4 a Verify proper connection of the adapter to the sensor b Verify that the airway adapter windows are clean and dry c If the adapter is properly connected and the...

Page 10: ...after its use When the previously used module is reconnected to the M Series unit module zeroing does not have to be repeated as long as no other LoFlo module has been connected and zeroed during the...

Page 11: ...lo module and ensure that it clicks into place 2 If the unit displays either of the following messages take the appropriate action 3 Place the airway adapter assembly at the proximal end of the airway...

Page 12: ...o module and rubber protective covering can be cleaned and disinfected by wiping with one or more of the following solutions 70 isopropyl alcohol 2 gluteraldehyde 10 bleach solution Use a soft cloth d...

Page 13: ...ftkey from the physiological monitoring menu cycles the display from the capnogram waveform to the plethysmograph waveform if SpO2 is installed to no second waveform displayed Use the Zoom softkey fro...

Page 14: ...airway circuit If the concentration of oxygen in the breathing circuit exceeds 60 and nitrous oxide is in use both O2 and N2O should be activated When the Comp softkey is pressed the unit displays the...

Page 15: ...alarm functions are disabled Background ECG analysis functions continue to operate as described in the AED section of the M Series Operator s Guide The ALARM SUSPEND button can be used to activate dea...

Page 16: ...ature unit displays WARM UP message 4 Perform a zero procedure if necessary see Zeroing the LoFlo CO2 Module on page 8 5 Breath normally into the adapter 6 Verify that the unit displays EtCO2 readings...

Page 17: ...3 kPa 0 95 mmHg OFF 0 12 5 OFF 0 12 6 kPa OFF High Respiration Rate Alarm Limit 120 respirations per min 5 150 respirations per min OFF Low Respiration Rate Alarm Limit 5 respirations per min 0 100 re...

Page 18: ...PU Oral Nasal CO2 sampling only pediatric Nasal CO2 Sampling with O2 Delivery Cannula Adult 8000 0356 SPU Nasal CO2 sampling with O2 delivery adult Nasal CO2 Sampling with O2 Delivery Cannula Pediatri...

Page 19: ...problem persists return for service CO2 DEVICE NOT READY The zero operation cannot be initiated because The sensor or module is still warming up No sensor or module is attached to the unit Wait for se...

Page 20: ...g the adapter type Clean airway adapter and reattach it If the problem persists or the adapter type was changed perform a mainstream airway adapter zero as described in Zeroing the Mainstream Sensor A...

Page 21: ...line is set below 0 mmHg Perform a sidestream module zero as described in Zeroing the LoFlo CO2 Module on page 8 CHECK CO2 LINE Sample line blockage Verify that the sample line is plugged into the mod...

Page 22: ...han 5 minutes Replace module ZERO CO2 MODULE Module cable connected to M Series unit for the first time Probe drift error detected or negative CO2 detected Perform module zero as described in Zeroing...

Page 23: ...lt pediatric adapters 50 ms Neonatal adapters 200 ms Respiration Rate Range 0 150 breaths per minute Respiration Rate Accuracy 1 breath per minute Sample Flow Rate N A 50 ml min Compensations Barometr...

Page 24: ...at maximum energy 360J or 1 5 hours minimum of continuous ECG monitoring or 1 0 hour of continuous ECG monitoring pacing at 60 mA 70 beats min For a new fully charged XL battery pack at 20 C 60 defibr...

Reviews: