End TIdal Carbon Dioxide (EtCO

2

)

EtCO

2

 - 7

 

2. Check that connections have been made correctly by 

verifying the presence of a proper CO

waveform on 

the M Series display. 

3. The sensor cable should face away from the patient. 

To secure the sensor cable safely out of the way, 
attach Sensor Cable Holding Clips to the airway 
tubing, then connect the sensor cable to the clips.

Applying an Airway Adapter with Mouthpiece

The disposable Pediatric/Adult airway adapter with 
mouthpiece can be used for spot checking CO

on 

non-intubated adult or pediatric patients.   

CAUTION!

The disposable Pediatric/Adult Airway 
Adapter with mouthpiece is intended for 
single patient use. Do NOT reuse or sterilize 
the adapter, as system performance will be 
compromised.

1. Remove adapter with mouthpiece from the package. 

Verify that the adapter and mouthpiece are intact and 
securely fastened to each other.

2. Attach the airway adapter to the CAPNOSTAT 3 

sensor, as follows:

a. Align the arrow on the bottom of the airway adapter 

with the arrow on the bottom of the 
CAPNOSTAT 3.

b. Press the sensor and airway adapter together until 

they click (see Figure 3).

3. If the unit displays the “CHECK CO2 ADAPTER” 

message, follow steps a through c, else go to step 4.

a. Verify proper connection of the adapter to the 

sensor.

b. Verify that the airway adapter windows are clean 

and dry.

c. If the adapter is properly connected, and the 

windows are clean and dry, then zero the adapter 
as described in "Zeroing the Mainstream Sensor/
Airway Adapter" on page 6.

4. Ensure the patient seals his or her mouth completely 

around the mouthpiece and breathes normally.

Note

A nose clip may be needed if the patient is exhaling 
through the nose. It is important that all, or most, of 
the exhalation be routed through the airway 
adapter.

Sidestream EtCO

2

 Setup

There are several steps involved with sidestream EtCO

2

 

setup, as follows:

Selecting a Sidestream Airway Adapter Kit or Cannula 

Attaching the LoFlo Module Cable and Inserting the 
Sample Cell

Zeroing the LoFlo CO2 Module

Applying a Sidestream Airway Adapter Kit 

Applying a Nasal or Nasal/Oral Cannula

Selecting a Sidestream Airway Adapter Kit

Select an airway adapter kit based on the patient's size, 
ET tube diameter, and monitoring situation. Airway 
adapter kits are disposable and single patient use. 

$ISPOSABLE.EONATAL!IRWAY!DAPTER

#!0./34!4#/

3ENSOR

4O0ATIENT 

6ENTILATOR7YE

Figure 2

Airway Adapter Kit

ET Tube Diameter

Adult/Pediatric Airway Adapter Kit

> 4.0 mm

Adult/Pediatric Airway Adapter Kit 
with Nafion® tubing

4O0ATIENT 

#!0./34!4#/

3ENSOR

-OUTHPIECE

!IRWAY!DAPTER

Figure 3

Summary of Contents for Capnostat 3

Page 1: ...9650 0223 01 Rev D TM End Tidal Carbon Dioxide EtCO2...

Page 2: ...l Corporation CAPNOSTAT is a registered trademark and LoFlo is a trademark of Respironics Novametrix LLC Cidex is a registered trademark of Advanced Sterilization Products a Johnson and Johnson Compan...

Page 3: ...mum filtration of fluids and contaminants and protecting the system from aspiration of these fluids In both systems the CO2 sensor generates infrared light and beams it through the airway adapter or s...

Page 4: ...es EtCO2 option Activate oxygen compensation if O2 levels in excess of 60 are introduced Activate N2O compensation if nitrous oxide is introduced into the airway circuit The presence of Desflurane bey...

Page 5: ...h windows in a vertical NOT a horizontal position This helps keep patient secretions from pooling on the windows Do NOT insert any object other than the sample cell into the sample cell receptacle on...

Page 6: ...t and property even if ZOLL has been advised of the possibility of such damages THE WARRANTY SET FORTH HEREIN IS EXCLUSIVE AND ZOLL MEDICAL CORPORATION EXPRESSLY DISCLAIMS ALL OTHER WARRANTIES WHETHER...

Page 7: ...EtCO2 setup as follows Attaching the CAPNOSTAT 3 sensor cable Selecting a mainstream airway adapter Attaching the airway adapter to the CAPNOSTAT 3 sensor Zeroing the mainstream sensor airway adapter...

Page 8: ...ximately 15 seconds The unit displays the message ZERO DONE upon completion of the zeroing Note Do not attempt zeroing for 20 seconds after removing the adapter from the patient s airway This time all...

Page 9: ...ow steps a through c else go to step 4 a Verify proper connection of the adapter to the sensor b Verify that the airway adapter windows are clean and dry c If the adapter is properly connected and the...

Page 10: ...after its use When the previously used module is reconnected to the M Series unit module zeroing does not have to be repeated as long as no other LoFlo module has been connected and zeroed during the...

Page 11: ...lo module and ensure that it clicks into place 2 If the unit displays either of the following messages take the appropriate action 3 Place the airway adapter assembly at the proximal end of the airway...

Page 12: ...o module and rubber protective covering can be cleaned and disinfected by wiping with one or more of the following solutions 70 isopropyl alcohol 2 gluteraldehyde 10 bleach solution Use a soft cloth d...

Page 13: ...ftkey from the physiological monitoring menu cycles the display from the capnogram waveform to the plethysmograph waveform if SpO2 is installed to no second waveform displayed Use the Zoom softkey fro...

Page 14: ...airway circuit If the concentration of oxygen in the breathing circuit exceeds 60 and nitrous oxide is in use both O2 and N2O should be activated When the Comp softkey is pressed the unit displays the...

Page 15: ...alarm functions are disabled Background ECG analysis functions continue to operate as described in the AED section of the M Series Operator s Guide The ALARM SUSPEND button can be used to activate dea...

Page 16: ...ature unit displays WARM UP message 4 Perform a zero procedure if necessary see Zeroing the LoFlo CO2 Module on page 8 5 Breath normally into the adapter 6 Verify that the unit displays EtCO2 readings...

Page 17: ...3 kPa 0 95 mmHg OFF 0 12 5 OFF 0 12 6 kPa OFF High Respiration Rate Alarm Limit 120 respirations per min 5 150 respirations per min OFF Low Respiration Rate Alarm Limit 5 respirations per min 0 100 re...

Page 18: ...PU Oral Nasal CO2 sampling only pediatric Nasal CO2 Sampling with O2 Delivery Cannula Adult 8000 0356 SPU Nasal CO2 sampling with O2 delivery adult Nasal CO2 Sampling with O2 Delivery Cannula Pediatri...

Page 19: ...problem persists return for service CO2 DEVICE NOT READY The zero operation cannot be initiated because The sensor or module is still warming up No sensor or module is attached to the unit Wait for se...

Page 20: ...g the adapter type Clean airway adapter and reattach it If the problem persists or the adapter type was changed perform a mainstream airway adapter zero as described in Zeroing the Mainstream Sensor A...

Page 21: ...line is set below 0 mmHg Perform a sidestream module zero as described in Zeroing the LoFlo CO2 Module on page 8 CHECK CO2 LINE Sample line blockage Verify that the sample line is plugged into the mod...

Page 22: ...han 5 minutes Replace module ZERO CO2 MODULE Module cable connected to M Series unit for the first time Probe drift error detected or negative CO2 detected Perform module zero as described in Zeroing...

Page 23: ...lt pediatric adapters 50 ms Neonatal adapters 200 ms Respiration Rate Range 0 150 breaths per minute Respiration Rate Accuracy 1 breath per minute Sample Flow Rate N A 50 ml min Compensations Barometr...

Page 24: ...at maximum energy 360J or 1 5 hours minimum of continuous ECG monitoring or 1 0 hour of continuous ECG monitoring pacing at 60 mA 70 beats min For a new fully charged XL battery pack at 20 C 60 defibr...

Reviews: