MAINSTREAM MESSAGES

Message/Symptom

Possible Cause(s)

Recommended Action(s)

Sensor not responding to 
breaths. CO

2

 waveform always 

at 0 mmHg with no messages 
displayed.

or

CO

2

 waveform baseline

> 25 mmHg with no messages 
displayed.

Sensor/Adapter was zeroed with a low level 
of residual CO

2

 in the adapter. CO

2

 

baseline is set below 0 mmHg.

Airway adapter type was changed with unit 
powered off. New sensor/adapter 
combination has not been zeroed.

Perform a mainstream airway adapter zero 
as described in "Zeroing the Mainstream 
Sensor/Airway Adapter" on page 6.

CHECK CO2 ADAPTER 
displayed during each 
inhalation. Message disappears 
during exhalation.

CHECK CO2 ADAPTER 
alternates with ZERO CO2 
ADAPTER shortly after 
breathing stops.

Sensor/Adapter was zeroed with a low level 
of residual CO

2

 in the adapter. CO

2

 

baseline is set below 0 mmHg.

Perform a mainstream airway adapter zero 
as described in "Zeroing the Mainstream 
Sensor/Airway Adapter" on page 6.

CHECK CO2 ADAPTER

Usually caused when the airway adapter is 
removed from the CAPNOSTAT 3, or when 
there is an optical blockage on the windows 
of the airway adapter. 
May also be caused by not having 
performed an adapter zero after changing 
the adapter type.

Clean airway adapter and reattach it. If the 
problem persists or the adapter type was 
changed, perform a mainstream airway 
adapter zero as described in "Zeroing the 
Mainstream Sensor/Airway Adapter" on 
page 6.

CHECK CO2 SENSOR

Sensor cable not properly plugged in or 
over temperature.

Check that sensor cable is plugged in and 
properly seated in the connector.
Check that sensor is not exposed to 
excessive heat.
If problem persists, replace sensor.

CO2 IN ADAPTER: WAIT

There is CO

2

 in the adapter when 

attempting to zero.
Zeroing was attempted within 20 seconds 
of previous zero operation.

Remove adapter from CO

2

 source including 

the patient’s - and your own - exhaled 
breaths, and ventilator exhaust valves.
Wait up to 20 seconds before retrying 
adapter zero.

REPLACE CO2 SENSOR

Sensor is defective or incompatible 
(message appears when “WARM UP’ 
message is displayed for more than 5 
minutes).

Replace sensor.

USE AIRWAY ADAPTER

CO

2

 sensor is on the “REF” or “0” cell when 

zeroing is started. 

Remove CO

2

 sensor from the “REF” or “0” 

cell and attach airway adapter to CO

2

 

sensor.

ZERO CO2 ADAPTER

Sensor cable connected to M Series unit for 
the first time. Zeroing error or probe drift 
error detected.

Perform a mainstream airway adapter zero 
as described in "Zeroing the Mainstream 
Sensor/Airway Adapter" on page 6.

ZEROING CO2 ADAPTER

Adapter zeroing in progress.

Wait for adapter zeroing to finish.

Summary of Contents for Capnostat 3

Page 1: ...9650 0223 01 Rev D TM End Tidal Carbon Dioxide EtCO2...

Page 2: ...l Corporation CAPNOSTAT is a registered trademark and LoFlo is a trademark of Respironics Novametrix LLC Cidex is a registered trademark of Advanced Sterilization Products a Johnson and Johnson Compan...

Page 3: ...mum filtration of fluids and contaminants and protecting the system from aspiration of these fluids In both systems the CO2 sensor generates infrared light and beams it through the airway adapter or s...

Page 4: ...es EtCO2 option Activate oxygen compensation if O2 levels in excess of 60 are introduced Activate N2O compensation if nitrous oxide is introduced into the airway circuit The presence of Desflurane bey...

Page 5: ...h windows in a vertical NOT a horizontal position This helps keep patient secretions from pooling on the windows Do NOT insert any object other than the sample cell into the sample cell receptacle on...

Page 6: ...t and property even if ZOLL has been advised of the possibility of such damages THE WARRANTY SET FORTH HEREIN IS EXCLUSIVE AND ZOLL MEDICAL CORPORATION EXPRESSLY DISCLAIMS ALL OTHER WARRANTIES WHETHER...

Page 7: ...EtCO2 setup as follows Attaching the CAPNOSTAT 3 sensor cable Selecting a mainstream airway adapter Attaching the airway adapter to the CAPNOSTAT 3 sensor Zeroing the mainstream sensor airway adapter...

Page 8: ...ximately 15 seconds The unit displays the message ZERO DONE upon completion of the zeroing Note Do not attempt zeroing for 20 seconds after removing the adapter from the patient s airway This time all...

Page 9: ...ow steps a through c else go to step 4 a Verify proper connection of the adapter to the sensor b Verify that the airway adapter windows are clean and dry c If the adapter is properly connected and the...

Page 10: ...after its use When the previously used module is reconnected to the M Series unit module zeroing does not have to be repeated as long as no other LoFlo module has been connected and zeroed during the...

Page 11: ...lo module and ensure that it clicks into place 2 If the unit displays either of the following messages take the appropriate action 3 Place the airway adapter assembly at the proximal end of the airway...

Page 12: ...o module and rubber protective covering can be cleaned and disinfected by wiping with one or more of the following solutions 70 isopropyl alcohol 2 gluteraldehyde 10 bleach solution Use a soft cloth d...

Page 13: ...ftkey from the physiological monitoring menu cycles the display from the capnogram waveform to the plethysmograph waveform if SpO2 is installed to no second waveform displayed Use the Zoom softkey fro...

Page 14: ...airway circuit If the concentration of oxygen in the breathing circuit exceeds 60 and nitrous oxide is in use both O2 and N2O should be activated When the Comp softkey is pressed the unit displays the...

Page 15: ...alarm functions are disabled Background ECG analysis functions continue to operate as described in the AED section of the M Series Operator s Guide The ALARM SUSPEND button can be used to activate dea...

Page 16: ...ature unit displays WARM UP message 4 Perform a zero procedure if necessary see Zeroing the LoFlo CO2 Module on page 8 5 Breath normally into the adapter 6 Verify that the unit displays EtCO2 readings...

Page 17: ...3 kPa 0 95 mmHg OFF 0 12 5 OFF 0 12 6 kPa OFF High Respiration Rate Alarm Limit 120 respirations per min 5 150 respirations per min OFF Low Respiration Rate Alarm Limit 5 respirations per min 0 100 re...

Page 18: ...PU Oral Nasal CO2 sampling only pediatric Nasal CO2 Sampling with O2 Delivery Cannula Adult 8000 0356 SPU Nasal CO2 sampling with O2 delivery adult Nasal CO2 Sampling with O2 Delivery Cannula Pediatri...

Page 19: ...problem persists return for service CO2 DEVICE NOT READY The zero operation cannot be initiated because The sensor or module is still warming up No sensor or module is attached to the unit Wait for se...

Page 20: ...g the adapter type Clean airway adapter and reattach it If the problem persists or the adapter type was changed perform a mainstream airway adapter zero as described in Zeroing the Mainstream Sensor A...

Page 21: ...line is set below 0 mmHg Perform a sidestream module zero as described in Zeroing the LoFlo CO2 Module on page 8 CHECK CO2 LINE Sample line blockage Verify that the sample line is plugged into the mod...

Page 22: ...han 5 minutes Replace module ZERO CO2 MODULE Module cable connected to M Series unit for the first time Probe drift error detected or negative CO2 detected Perform module zero as described in Zeroing...

Page 23: ...lt pediatric adapters 50 ms Neonatal adapters 200 ms Respiration Rate Range 0 150 breaths per minute Respiration Rate Accuracy 1 breath per minute Sample Flow Rate N A 50 ml min Compensations Barometr...

Page 24: ...at maximum energy 360J or 1 5 hours minimum of continuous ECG monitoring or 1 0 hour of continuous ECG monitoring pacing at 60 mA 70 beats min For a new fully charged XL battery pack at 20 C 60 defibr...

Reviews: