ZOLL AED PLUS Administrator’s Guide
7
PEDIATRIC PADS.
ZOLL AED Plus has detected pediatric electrode pads connected
to it and adjusted defibrillation energy settings to pediatric levels.
DON’T TOUCH PATIENT,
ANALYZING.
Do not touch patient, an ECG rhythm analysis is in progress or
about to begin.
SHOCK ADVISED.
ECG rhythm analysis has detected the presence of VF or
shockable VT.
NO SHOCK ADVISED.
ECG rhythm analysis has detected a rhythm that is not treatable
by defibrillation.
ANALYSIS HALTED. KEEP
PATIENT STILL.
ECG rhythm analysis has been halted due to excessive ECG
signal artifact. Stop any ongoing CPR and keep the patient as
motionless as possible.
DON’T TOUCH PATIENT.
PRESS FLASHING SHOCK
BUTTON.
Warn all persons in attendance of the patient to stand clear and
stop touching the patient. Press the Shock Button to deliver
defibrillation therapy.
RELEASE SHOCK BUTTON.
Shock Button was depressed before the defibrillator was ready to
defibrillate. Release the Shock Button and press again after the
ready tone sounds.
SHOCK DELIVERED.
A defibrillation shock has just been delivered to the patient.
NO SHOCK DELIVERED.
No shock was delivered to the patient because rescuer failed to
press Shock Button or an error condition was detected.
n SHOCKS DELIVERED.
A total of
n
shocks have been delivered since the ZOLL AED
Plus was turned on.
START CPR.
Begin CPR.
CONTINUE CPR.
Continue providing CPR. This prompt may also be issued if the
ZOLL AED Plus CPR monitoring function fails to detect chest
compressions at least ¾ of an inch deep.
PUSH HARDER.
CPR compressions are consistently less than 1 ½ inches deep.
GOOD COMPRESSIONS.
After prompting to Push Harder, rescuer has succeeded in
delivering chest compressions at least 1 ½ inches deep.
STOP CPR.
Stop CPR, the ZOLL AED Plus is about to begin an ECG rhythm
analysis.
Voice Prompt
Definition