28
ZOLL AED PLUS Administrator’s Guide
DEFIBRILLATOR (cont’d)
CPR
*Metronome Rate: Variable 60 to 100 CPM
Depth: ¾" to 3"; 1.9 to 7.6 cm
Defibrillation Advisory
Evaluates electrode connection and patient ECG to determine if
defibrillation is required.
Shockable Rhythms: Venticular fibrillation with average
amplitude>100 microvolts and wide complex ventricular
tachycardia with rates greater than 150 BPM (adult mode) and
200 BPM (pediatric mode). Refer to ECG Analysis Algorithm
Accuracy Section for sensitivity and specificity performance.
Electrode Patient Impedance
Measurement Range
0 to 300 ohms
Defibrillator Electrode ECG
Circuitry
Protected
ECG Bandwidth
2-30Hz
Display Format
Optional LCD with Moving Bar
Size: 2.6" x 1.3"; 6.6 cm x 3.3 cm
Viewing Time: 2.6 seconds
Display Sweep Speed
25 mm/sec
Battery Capacity
Typical new battery (20° C):
• stand-by life with batteries installed: 5 years, or
• 300 continuous shocks, or
• at least 13 hours of continuous Monitoring time
PC Minimum Requirements
Windows
®
98, Windows
®
2000 Windows
®
NT, Windows
®
XP
IBM-compatible PII with 16550 UART (or higher) computer
64MB RAM.
VGA monitor or better
CD-ROM drive
IrDA
™
port
20MB disk space
*
Testing reports validating performance and accuracy of CPR depth measurement capability,
metronome feature function and rescuer performance, and the PASS (Passive Airway Support
System) cover function are on file with ZOLL Medical Corporation and are available for review.
Contact ZOLL Technical Support to request a copy of the following report(s) if desired:
• Using the ZOLL AED Plus Cover to Aid in Airway Patency
• Depth and Compression Rate Response of the ZOLL AED Plus CPR System
• ZOLL AED Plus CPR System Test Results.