Out of Range Control Solution Results
•
Repeat the test if control solution results are outside the range printed on
the test strip vial .
•
Repeat the test to confirm the results .
•
The correct test procedures may not have been followed .
•
Stop using the meter if control solution results are consistently outside the
range printed on the test strip vial . Contact Customer Care .
Causes of out-of-range results include:
•
Expired or bad control solution
•
Expired or bad test strip
•
Testing error
•
Watered-down control solution
•
Meter malfunction
•
Test strip deterioration
•
Control solution test done outside 59° – 104° F (15° – 40° C)
•
Correct code is not set in the meter
Performing an AlphaTRAK Control Solution Test
1.
Check control solution expiry date .
2.
Remove test strip from vial .
3.
Insert test strip to turn meter on .
Note:
The meter turns off after 2 minutes of inactivity .
Remove and reinsert the unused test strip to restart
the meter .
If the meter does not turn on go to Troubleshooting section .
IMPORTANT:
Do
not
use control solution past expiry date . Discard
control solution 3 months after opening or on the expiry date printed
on the bottle whichever comes first . (Example: open April 15, discard
July 15; write the discard date on the side of the bottle .)
Top
Printed
side up
Sample
areas
Contact Customer Care at 1-888-963-8471