![Zoetis AlphaTRAK 2 Meter User Manual Download Page 18](http://html.mh-extra.com/html/zoetis/alphatrak-2-meter/alphatrak-2-meter_user-manual_936832018.webp)
English
English
29
30
8.
Compare the control solution result to the range printed on the test strip vial or
package insert . The result should fall within the range .
Note:
If you see an “
Er1, Er2, Er3, Er4 or Er6
” on the display, it is an error code . Go
to Error Codes section .
≈ 7.7 mmol/L
(Example only)
9.
Press and hold
c
for
2
seconds until
appears on the display to mark
the result as a control solution result . To unmark a control solution test result,
press and hold
c
for
2
seconds . A value with the
indicates you have
marked this reading as a control solution test .
10.
Remove the test strip to turn meter off .
Using Meter Memory
Purpose
The purpose of this section is to review your stored test results and 7-, 14- and 30-day averages .
Viewing Blood Glucose Averages in Memory
Start with the meter off (no test strip inserted) . Press
m
. The 7-day average
appears (control solution, LO and HI readings
not
included) .
Note:
To exit meter memory at any time, press and hold
m
.
Example:
Memory Display
the average
n= number of results included in average
•
Press
c
to scroll through 14- and 30-day averages and all stored test
results . The most recent result appears first .
•
Press
m
to scroll backwards .
Contact Customer Care at 1-888-963-8471