Application screen
4
21
Energy / Bar graph
Shows the selected pulse energy. When application is active the bar
graph is filled in.
Setting the pulse energy can be done either before or during treatment,
using the left controller.
The pulse energy (on the applicator) can be set from 60 mJ to 185 mJ,
adjustable in 10 mJ steps (at 16 Hz max. 120 mJ, at 22 Hz max. 90 mJ).
Save
After changing the parameters based on individual needs press the
button
′
Save
′
for saving the settings either in the Favorites list or the
Memory list.
Count direction
Press to set the count direction (increasing or decreasing) of the number
of pulses.
Number of pulses
Shows the preselected pulse numbers as well as the current number of
pulses delivered to the patient.
Also the count direction (decreasing in this case) is shown.
Pressing the 'Number of pulses' field opens the Input window, defining
preselection and count direction.
Note:
Z
Wave offers two options for pulse delivery:
Pulse delivery without
preselecting the pulse
number
For pulse delivery with no preselected number of pulses, the application
is not ended by the device. As long as the footswitch is activated, pulses
will be delivered.
For pulse delivery with no preselected number, only the upwards count
direction is active.