Manufacturer’s declaration of
Electromagnetic Compatibility
15
45
Medical electrical devices such as Z
Wave
are subject to special precautions with regard to electromagnetic compatibility
(EMC) and must be installed and commissioned in accordance with the EMC advice given in the instructions for use and
accompanying documents.
Portable and mobile RF communication systems (e.g. mobile phones) may interfere with medical electrical equipment.
Z
Wave
should only be operated with the original mains cable specified in the list of contents delivered.
Operating the device with any other mains cable can lead to increased emissions or reduced interference immunity of the
device.
Guidelines and manufacturer's declaration – electromagnetic interference
The device Z
Wave
is intended for operation in an electromagnetic environment as indicated below. The customer or
user of the Z
Wave
should ensure that it is operated in such an environment.
Interference tests
Conformity
Electromagnetic environment guideline
RF emissions according to CISPR 11
Group 1
The device Z
Wave
uses RF energy solely for
its internal functioning. Its RF emission is
therefore very low and it is unlikely that this
will cause interference to neighbouring
electronic equipment.
RF emissions according to CISPR 11
Class A
The device Z
Wave
is suitable for use in all
installations including those in a residential
environment and those which are directly
connected to the public mains network which
also supplies buildings which are used for
residential purposes.
Harmonic emissions according to IEC
61000-3-2
Class A
Voltage fluctuation emissions and flicker
according to IEC 61000-3-3
Conforms
The device should not be used when placed immediately next to or stacked on top of other devices. If operation is necessary
when immediately next to or stacked on top of other devices, the device should be monitored to ensure it is operating as
intended in this arrangement.