Function Test
16
Page 34
Soleo Sono performs a self-test when switched on to check all the internal
components.
In the event of a fault, an error message appears.
In addition, an extended function test can be performed as described below.
These tests should be performed monthly or in case of any doubt regarding the
correct function of the device.
Ultrasound
Select the ultrasound head and cover the ultrasound head surface with coupling gel.
In the case of starting therapy with low power, the coupling display must show more
than 90%.
Run the test with both ultrasound heads, one after the other.
Then clean the ultrasound heads.