Application Information
3.1 General
3
Page 4
Prior to the application of the device to the patient, the user should become
acquainted with the operating instructions or the individual treatment methods and
also with the indications / contraindications, warnings and directions for use.
Additional sources of information on ultrasound therapy should also be taken into
consideration.
This user manual should always be kept with the device to enable the persons
charged with the operation of the device to access it at any time.
Caution
!
To avoid the risk of electric shock, the device must be disconnected from the mains
supply before performing any maintenance or cleaning activities.
Caution
!
Before use, ensure that the device is powered via a properly earthed plug with a
grounded outlet (electrical installation according to DIN VDE 0100 Part 710).
The
device must only be operated with the supplied power cord. The power cord must be
protected against mechanical stress.
Caution
!
Magnetic and electric fields can influence the function of the device. Therefore, the
device must not be operated in the vicinity of devices that generate strong
electromagnetic fields (X-ray or diathermy devices, MRI scanners). Please keep a
safe distance of several metres.
Caution
!
The device is not suitable for use in areas with an explosive, combustible or
combustion-promoting atmosphere.
Caution
!
During use, the device must be placed in such a way that direct access to the
device's central power supply is possible, so that it can be disconnected from the
mains at any time.
Inspect the device before use. If damaged, it must not be used.
Caution
!
Only accessories provided by Zimmer MedizinSysteme GmbH are to be used and no
other devices must be connected.
Caution
!
The device may cause malfunctions or may interfere with the operation of equipment
in the vicinity. It may be necessary to take appropriate remedial action, such as the
realignment or re-arrangement of the device or the shield.
Caution
!
No change to the device or the medical system may be carried out during the product
life of the device.
The entire medical system is suitable for application in the patient environment.
Caution
!
To safely disconnect the unit from the mains supply, remove the mains plug from the
socket housing or socket.