14
SECTION
7
Implant Femoral Components
Persona
Trabecular Metal
PS Femoral Component
Size
7
Right
42-5008-062-02
Persona
Trabecular Metal
CR Femoral Component
Size
7
Right
42-5028-062-02
1. Unlock Collar
2. Insert
3. Release
4. Rotate
Implant Femoral Components
After the implants have been chosen, make a final check to
ensure that the size chosen for the femoral, tibial, and articular
surface are compatible.
Cementless Implantation Option
•
With the knee in 70°-90° of flexion, retract the soft tissue in
the desired manner. Attach the Femoral Inserter/Extractor to
the femoral component
(Fig. 39a & 39b)
. Insert the femoral
component onto the distal femur by translating the component
laterally until the lateral peg aligns with the drill hole in the
lateral femoral condyle. Take care to avoid scratching the
implant component surfaces. After the femoral component is
placed on the femur and the inserter/extractor is removed, the
impaction head/inserter assembly can be used to impact the
component onto the femur. If this insertion/impaction method
is used,the inserter handle must be in the closed and locked
position. Ensure that soft tissue is not trapped beneath the
implant. Alternatively, assemble the Quick Connect Handle to
the Femoral Impactor Head
(Fig. 40)
. Use the Femoral Impactor
Head assembly to fully seat the femoral component
(Fig. 41)
.
Remove retractors, and check the medial and lateral sides to
make sure the femoral component is fully impacted distally.
Fig. 39a
Fig. 39b
Fig. 41
Fig. 42
Fig. 40
Quick Connect
Handle
00-5901-034-00
Femoral Impactor
Head
00-5901-032-00
Persona
Femoral
Inserter/Extractor
42-5099-092-00
Persona
Femoral
PS Impactor Pad
42-5099-093-00
Persona
Femoral
CR Impactor Pad
42-5099-094-00
TECHNIQUE TIP
7.A
- Apply upward force on the Femoral Inserter/Extractor while
impacting the femoral implant to ensure component is placed
in proper orientation.
- Carefully engage the pegs or the intercondylar box of the femoral
implant into the pre-drilled peg holes or resected box and impact the
femur onto the bone until stable. Periodically check the alignment
with the bone cuts to ensure the implant is seating properly.
- The Trabecular Metal material extends above the femoral implant
rail by approximately 0.5mm and has a dark color, which may appear
not fully seated. If no light is visible beneath the distal surface
and the bone, the implant is fully seated. Small gaps between the
Trabecular Metal surface and the chamfer bone cuts may be visible
and are acceptable
(Fig. 42)
.
- Do not intra-operatively re-insert a previously extracted Persona
Trabecular Metal femoral implant.