10
SECTION
5
CR Femoral Finishing
Pinch to
release
•
Remove any posterior osteophytes or overhanging bone
on the femur to facilitate maximum knee flexion. Attach
the Femoral Inserter/Extractor to the correct CR Femoral
Provisional by inserting the tab on the Femoral Inserter/
Extractor arm into the anterior notch in the CR Femoral
Provisional and close the handle on the Femoral Inserter/
Extractor to secure the CR Femoral Provisional
(Fig. 26)
.
Place the correct CR Femoral Provisional onto the femur in
the desired medial/lateral position. Impact the end of the
Femoral Inserter/Extractor handle and apply an upward
force to fully seat the CR Femoral Provisional onto the femur
in the proper orientation.
(Fig. 27)
. To remove the Femoral
Inserter/Extractor from the CR Femoral Provisional, pinch
the release lever on the proximal portion of the Femoral
Inserter/Extractor
(Fig. 27)
. Alternatively, if the CR Femoral
Provisional is placed on the femur by hand, the CR femoral
Impaction Pad/inserter assembly can be used to impact
the provisional onto the femur. If this insertion/impaction
method is used, the inserter handle must be in the closed
and locked position. For additional fixation of the fully
seated provisional, insert the 25mm x 3.2mm screw
(2.5mm female hex) with the 2.5mm Male Hex Driver through
the hole in the lateral anterior flange of the CR Femoral
Provisional
(Fig. 28)
.
Note: Do NOT use a 48mm screw in the anterior lateral fixation
hole. 48mm screws are not recommended due to potential
bone perforation.
•
Once the tibia has been prepared, a trial range of motion can
be performed to assure proper positioning of the CR Femoral
Provisional prior to femoral peg hole preparation. Once
desired medial-lateral placement has been attained, drill the
peg holes for the femoral implant through the CR Femoral
Provisional with the NexGen
®
6.4mm Patella/Femoral Drill
(Fig. 29)
. If a screw was used to provide adjunct fixation,
remove the screw from the anterior flange in the CR Femoral
Provisional. The Slaphammer can be used to remove the
CR Femoral Provisional from the bone using the medial
hole on the CR Femoral Provisional
(Fig. 30)
. Alternatively,
the Femoral Inserter/Extractor can be re-attached to the CR
Femoral Provisional to remove it from the bone. If necessary,
place the round end of the Slaphammer in the extraction
hole of the Femoral Inserter/Extractor to facilitate removal.
TECHNIQUE TIP
5.C
Do not impact the anterior flange of the CR Femoral Provisional
for removal, as this may damage the CR Femoral Provisional.
Fig. 26
Fig. 27
Fig. 28
Fig. 29
Fig. 30
Persona
CR Femoral
Provisional
Size
7
Right
42-5027-062-02
Persona
Femoral
CR Impactor Pad
42-5099-094-00
Persona
Femoral
Inserter/Extractor
42-5099-092-00
NexGen
6
.
4
mm
Patella/Femoral
Drill
00-5120-052-01
Persona
Slaphammer
42-5099-037-00
1. Insert
2. Rotate
3. Extract
2
.
5
mm Male
Hex Driver
42-5099-025-00
25
mm x
3
.
2
mm
(
2
.
5
mm female hex)
42-5099-025-25
Instruments