5
SECTION
3
Side Designation
2
.
5
mm Male
Hex Driver
42-5099-025-00
Persona
Anterior
Referencing Sizer
42-5099-088-10
•
Apply the Anterior Referencing Sizer so that the flat surface
of the Anterior Referencing Sizer is flush against the resected
surface of the distal femur and the feet of the Anterior
Referencing Sizer are flush against the posterior condyles.
Center the Anterior Referencing Sizer mediolaterally. Both the
vertical and horizontal portions of the Anterior Referencing
Sizer provide visual cues relative to the AP and epicondylar
axes of the femur to help ensure that desired external rotation is
attained. If the 3° external rotation holes are to be used to set
external rotation, the etched line on the Anterior Referencing
Sizer should be positioned so it is in line with Whiteside’s
line
(Fig. 14)
to optimize the M/L position of the drill holes for
subsequent Anterior Referencing 4-in-1 Cut Guide placement.
If the 5° external rotation holes are to be used, the sizer can
be positioned with the etched line on the Anterior Referencing
Sizer 4mm laterally from Whiteside’s line to better center the
drill holes for subsequent Anterior Referencing 4-in-1 Cut
Guide placement, due to the M/L offset of the holes. Hold the
Anterior Referencing Sizer in place and if necessary, secure the
Anterior Referencing Sizer to the femur using 25mm x 3.2mm
(2.5mm female hex) screws
(Fig. 15)
in one or both of the
holes on the lower portion of the Anterior Referencing Sizer to
help draw the Anterior Referencing Sizer adjacent to the distal
femur, particularly in MIS situations.
Note: Do not use 48mm screws for fixation of the Anterior
Referencing Sizer. 48mm screws are not recommended due to
potential bone perforation.
Size Femur and Establish External Rotation
Size Femur and Establish
External Rotation
•
Rotate the feet of the Anterior Referencing Sizer so the
appropriate “Left” or “Right” markings are visible as the
Anterior Referencing Sizer is placed on the bone
(Fig. 13)
.
External rotation can be set at 3° or 5° from the posterior
condylar axis.
TECHNIQUE TIP
3.A
Remove any osteophytes that interfere with instrument positioning.
TECHNIQUE TIP
3.B
Do not impact the Anterior Referencing Sizer onto the femur.
Fig. 14
Fig. 15
Fig. 13
Instruments
25
mm x
3
.
2
mm
(
2
.
5
mm female hex)
42-5099-025-25