11
Instruments
SECTION
6
Persona
PS
Femoral Provisional
Size
7
Right
42-5007-062-02
Persona
Femoral
PS Impactor Pad
42-5099-093-00
Fig. 31a
Fig. 31b
•
Assemble the Femoral PS Impactor Pad to the Femoral
Inserter/Extractor. Hold the Femoral Inserter/Extractor,
with the handle in the open position and insert the Femoral
PS Impactor Pad, aligning the "PS" on the Femoral PS
Impactor Pad with the arrow on the Femoral Inserter/
Extractor
(Fig. 31a)
. The Femoral PS Impactor Pad is keyed,
so the Femoral PS Impactor Pad may have to be rotated
while placing and aligning the Femoral PS Impactor Pad
onto the Femoral Inserter/Extractor .
•
Femoral sizes are provided in two profiles, Standard
and Narrow. The Standard femoral provisionals have
intermittent cutouts around the periphery, with the inner
dimension representing the outer profile of the Narrow
femoral implant and the outer dimension representing
the outer profile of the Standard femoral implant
(Fig. 31b)
.
Care should be taken to use the appropriate Standard or
Narrow implant as is related to side (left or right) and size
based on the provisional fit and ROM provided during the
trialing phase.
PS Femoral Finishing
and PS Box Preparation
TECHNIQUE TIP
6.A
Reference the orientation and size etched and/or engraved markings to
identify the correct provisional.
Persona
Femoral
Inserter/Extractor
42-5099-092-00
Narrow Cutouts
Alignment Arrows
Release Lever
PS Femoral Finishing and PS Box Preparation
Note: Narrow phantom
lines are for illustration
clarity only.