A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
29
Table 3.2 Troubleshooting
SYMPTOM
POSSIBLE CAUSES
1. Main Cuff or Second Cuff will not inflate.
a) Membrane Panel not properly plugged into P8.
b) Tubing inside unit may be pinched or
improperly connected.
c) Deflate valve is stuck open.
d) Pump not properly plugged into P14.
e) Pump’s electrical harness damaged.
f) INFLATE button not working.
g) Valve’s electrical harness damaged.
h) Defective valve driver circuitry.
2. Main Cuff or Second Cuff will not deflate.
a) Membrane Panel not properly plugged into P8.
b) DEFLATE button not pressed long enough
(at least 2 seconds).
c) Deflate valve is stuck shut.
d) DEFLATE button not working.
e) Valve’s electrical harness damaged.
f) Defective valve driver circuitry.
3. No green AC Indicator light.
a) Unit not plugged into wall outlet.
b) No power at wall outlet.
c) Mains AC harness not properly plugged into P2.
d) Blown fuse(s).
e) Membrane Panel not properly plugged into P8.
f) Defective
AC
indicator.
g) Defective AC indicator circuitry.
4. No flashing orange Battery Indicator light.
a) Unit running on AC.
b) Membrane Panel not properly plugged into P8.
c) Defective battery indicator.
d) Defective battery indicator circuitry.
5. ALARM SILENCE button not working.
a) ALARM SILENCE button not properly plugged
into P1.
b) Non-silenceable alarm (System Failure).
c) ALARM SILENCE button defective.
d) Defective alarm silence circuitry.
6. Alarm indicator light not working.
a) Membrane Panel not properly plugged into P8.
b) Defective alarm indicator.
c) Defective alarm indicator circuitry.