A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
Table 2.4 EMC Guidance and Declaration - EMC Emissions
Guidance and manufacturer’s declaration – electromagnetic emissions
The
A.T.S.
3000 is intended for use in the electromagnetic environments specified below. The customer or the user of the
A.T.S.
3000 should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The
A.T.S.
3000 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The
A.T.S.
3000 is suitable for use in all establishments other than
domestic and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
Flicker emissions
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration – electromagnetic immunity
The
A.T.S.
3000 is intended for use in the electromagnetic environments specified below. The customer or the user of the
A.T.S.
3000 should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment – guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact (230 Vac
50 Hz / battery)
± 8 kV air (230 Vac/
battery)
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic materials, the
relative humidity should be at least 30%
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/output
lines
± 2 kV for power supply
lines (120 Vac 60 Hz / 230
Vac 50 Hz)
± 1 kV for input/output
lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
(120 Vac 60 Hz / 230 Vac
50 Hz)
± 2 kV line(s) to earth
(120 Vac 60 Hz / 230 Vac
50 Hz)
Mains power quality should be that of a typical
commercial or hospital environment.
In the event of a power surge, the
A.T.S.
3000’s
LCD screen may malfunction, but all other
functions should continue to operate. Power
down the device to reset. Consult the Operator and
Service Manual for additional specifications.
Table 2.5 EMC Guidance and Declaration - EMC Immunity/Disturbances